D620.388.00_QS_UK_01
29 / 42
Indirect mode - 1 MPa range
This configuration is used for indirect mode at high pressure range.
Connections
Connection Option / description
Air supply to1
Connection of the regulator air supply to the filter input (1.2 MPa)
4 to3
Connection of the regulator output to the part under test
3 to2
Connection of the part to the test input (indirect mode option)
5 to ATM
Connector left free to the atmosphere
Air supply to6
Connection of the air supply to the valves pilot input (0.6 MPa)