Error
Error display is released automatically when it is in the measured and display range.
  
HHHH
Disconnect power supply 
and check the cables.
LLLL
D-HH
Reset within the display 
range. 
D-LL
F-HH
-PF-H
PF-L
OVER
Reset within the zero range. 
Reset
01. 
alternately for 0.5 sec in turn. 
02. 
03. 
to run mode. 
•  The below describes based on OUT1. 
•  OUT1 and OUT2of output operations are same. It operates individually by the set output 
operation mode. 
•  When changing output operation mode,  
are reset.
MODE 
 OFF
OFF
OUT
output
HIGH
Hys
OUT
LOW
Hys
OUT
HL
Hys
OUT
 
HL-G
Hys
OUT
• 
• 
transducer. 
- LEAD: current phase leads voltage phase 
• It is available to accept several outputs of the power factor transducer by high / low-
limit input value analog output value setting in the power factor transducer.
• 
•  Setting range: from min. to max. selected value from input range
•  E.g.:  When setting 200V in input range, high / low-limit input value are available to set from 
-200.0 to 200.0. When setting 20V, high / low-limit input value are available to set from 
-20.00 to 20.00.  
value
cosØ
value
1
0.5
LEAD 
2
Default
H-RG L-RG
 500.0 000.0
 5.000 0.000
•  E.g. 1: When the output of the power factor transducer is DC 4 - 20 mA
1. Connect the output to the input terminal 7 (+), 8 (-) of this unit, then set input range as 4-20.
2.  When setting the input range as 4-20, low-limit input value is set as 4.00 and high-limit input 
value is set as 20.00 automatically. 
20 mA, it displays 0.50.
 
 
1-5B.
2.  Set high-limit input value as 5.00 and low-limit input value as 1.00 for the output of the 
power factor transducer.
V, it displays 0.50.
1V
5V
0.50
0.75
0.90
1.00
-0.90
-0.75
-0.50
Low
1
0.9
0.5
LEAD 
•  
1. Set OUT1 output operation mode as HL at parameter 2 group.
2.  Set OUT1 high-limit output setting value as 0.90 and OUT1 low-limit output setting value as 
-0.90 at parameter 0 group.
LEAD
Hys
0.90
HL
OUT1 high-limit 
output setting value
1.00
-0.90
Hys