PARAMETER X/XY REMARK DEFAULT
P8.x
Division Select
0-2 0 = 1 1 = 2 2 = 5 2
P9.x
Decimal Point in Calibraon
0-5
0 = x1
1 = x0.1
2 = x0.01
3 = x0.001
4 = x0.0001
5 = 10 /19 /15
2
P10.x
Calibraon Unit
0-1
0 = kg
1 = lb
0
P11.x
Weighing Units Enabled
0-6
0 = only kg
1 = only lb
2 = only lb:oz
3 = kg or lb
4 = kg or lb:oz
5 = lb or lb:oz
6 = kg, lb, or lb:oz
3
P12.x
Power-on Zero-point Range
0-7
0 = Calibraon zero-point ±1%FS
1 = Calibraon zero-point ±2%FS
2 = Calibraon zero-point ±5%FS
3 = Calibraon zero-point ±10%FS
4 = Calibraon zero-point ±20%FS
5 = Calibraon zero-point ±50%FS
6 = Calibraon zero-point ±100%FS
7 = No limitaon
7
P13.x
Zero Range for ZERO Key
0-9
0 = Power-on zero-point ±1%FS
1 = Power-on zero-point ±2%FS
2 = Power-on zero-point ±3%FS
3 = Power-on zero-point ±4%FS
4 = Power-on zero-point ±5%FS
5 = Power-on zero-point ±10%FS
6 = Power-on zero-point ±20%FS
7 = Power-on zero-point ±50%FS
8 = Power-on zero-point ±100%FS
9 = No limitaon
6
P14.x
Data as Current Power-on
Zero Point When Weight
Signal IS IN Power-on
Zero-point Range
0-2
0 = Current weight
1 = Calibraon zero-point
2 = Switch-o zero-point
0
P15.x
Data as Current Power-on
Zero Point When Weight
Signal IS NOT IN Power-on
Zero-point Range
0-3
0 = Current weight
1 = Calibraon zero-point
2 = Switch-o zero-point
3 = Connuous display
3
P16.x
Zero Tracking Range
0-8
0 = 0d, No Tracking
1 = ±0.25d 2 = ±0.5d 3 = ±1d 4 = ±1.5d
5 = ±2d 6 = ±3d 7 = ±4d 8 = ±5d
6
P17.x
Data Filter Intensity
0-3 0 = Very Weak 1 = Weak 2 = Middle 3 = Strong 2