9.
PARAMETER X/XY REMARK DEFAULT
P1.xy
Auto-o Time
00-15
00 = No Auto-O
01-15 = Minutes Auto-O Time
05
P2.xy
Hold/Print Funcon
00-02
00 = Only Hold Funcon
01 = Only Print Funcon
02 = Both Funcons (<3 sec = Print, >3 sec = Hold)
02
P3.xy
Hold Funcon
0-50
0 = Disable Hold Funcon
1 = Hold Larger Weight Reading
2-50 =
Hold reading when variety is within ±2-50d,
auto release Hold funcon when weight is <10d
and auto-hold new stable weight (> 10d)
02
P4.x
RS232 & USB Funcon
0-7
0 = No RS232 and USB
1 = Press PRINT key to output display data when
scale is stable
2 = Press PRINT key to output gross, tare, and net
weight when scale becomes stable
3 = Connuously outputs display data
4 = Connuously outputs gross, tare, and net weight
5 = Outputs display data one me when scale is stable
6 = Outputs gross, tare, and net weight one me
when scale becomes stable
7 = Bi-direconal communicaon
2
P5.x
Communicaon Baud Rate
0-4
0 = 1200bps
1 = 2400bps
2 = 4800bps
3 = 9600bps
4 = 19200bps
3
P6.x
Communicaon Format
0-2 0 = 8N1 1 = 7O1 2 = 7E1 0
P7.xy
Resoluon
00-32
00 = 500 08 = 2400 16 = 7500 24 = 35000
01 = 600 09 = 2500 17 = 8000 25 = 40000
02 = 750 10 = 3000 18 = 10000 26 = 50000
03 = 800 11 = 3500 19 = 12000 27 = 60000
04 = 1000 12 = 4000 20 = 15000 28 = 70000
05 = 1200 13 = 5000 21 = 20000 29 = 75000
06 = 1500 14 = 6000 22 = 25000 30 = 80000
07 = 2000 15 = 7000 23 = 30000 31 = 100000
32 = Factory Preset N (0<N<1000000)
10
USER PARAMETERS
1. When the scale is in normal weighing mode, press and hold the ZERO and UNIT/COUNT keys
for 3 seconds until is shown on the display.
2. In the Setup mode, press the ↑ key to change the flashing digits, press the → key to shift the
flashing data entry position from left to right, and press the TARE key to confirm and move to
the next parameter setting. Press the ZERO key to exit the Setup mode.
PARAMETERS SETTING SUMMARY