Azbil Corporation Installation
Model PTG60/70/71/72 - Smart Pressure Transmitter 2-1
Chapter 2 : Installation
Introduction
7KLVVHFWLRQGHVFULEHVYDULRXVLQVWDOODWLRQPHWKRGVIRUGL൵HUHQWW\SHVRISURFHVV
interface.
2-1 : Air inlet
$0RGHO37*LVDJDXJHSUHVVXUHWUDQVPLWWHU7KHEXLOWLQVHPLFRQGXFWRUSUHVVXUH
sensor measures atmospheric pressure as the reference pressure.
Install this device in such a way that water will not enter the air inlet and powder or
viscous materials will not be deposited on it.
,IWKLVGHYLFHZLOOEHLQVWDOOHGXSVLGHGRZQDWWKHERWWRPRIWKHWDQNZLWKWKHSURFHVV
FRQQHFWLRQSRUWIDFLQJXSZDUGRULISRZGHURUYLVFRXVPDWHULDOPD\DGKHUHWRRU
DFFXPXODWHRQWKHGHYLFH¶VH[WHULRULQVHUWWKHFKORURSUHQHUXEEHU¿WWLQJLQFOXGHG
ZLWKWKHSURGXFWLQWRWKHDLULQOHW
,IWKLVGHYLFHZLOOEHH[SRVHGWRKHDY\VSODVKLQJRULIDJUHDWDPRXQWRISRZGHURU
YLVFRXVPDWHULDOZLOODGKHUHWRRUDFFXPXODWHDURXQGLWLWLVDOVRSRVVLEOHWRLQVHUWD
PPGLDPHWHUWXEHLQWRWKLVLQOHWWRLQWURGXFHDLUDWDGL൵HUHQWSRVLWLRQ
Air inlet
Fitting for air inlet
(chloroprene rubber, included with the product)
When the air inlet tting is attached
Figure 2-1 $LULQOHWDQGUXEEHU¿WWLQJ
)RUPRGHOVZLWKDQLQWHUQDODLULQOHWRSWLRQDOWKHLQOHWLVORFDWHGLQWKHFDEOHZLULQJ
VSDFHVRDVQRWWREHH[SRVHGWRWKHRXWVLGH