Installation Azbil Corporation
2-14 Model PTG60/70/71/72 - Smart Pressure Transmitter
Handling precautions
• Weld or bind with a sealing tape securely to the piping joint to prevent a process
OHDNDJHIURPRFFXUULQJ
• Avoid installing the Transmitter with its front side facing down. Doing so invites
water, dirt and other sediment to deposit at the wetted part, causing an inaccurate
measurement.
• ,IWKHSURFHVVWHPSHUDWXUHH[FHHGV&)WDNHVRPHPHDVXUHVVXFKDV
installing a siphon, to defuse the high temperature at the wetted part so it does not
H[FHHG&)
• Install the Transmitter in a location where liquid process does not freeze.
2WKHUZLVHWDNHVRPHPHDVXUHVWRNHHSWKHSURFHVVÀXLGIURPIUHH]LQJ
• ,QVWDOOWKH7UDQVPLWWHULQDORFDWLRQZKHUHWKHUHLVQRH[FHVVLYHLPSDFWYLEUDWLRQ
RUSUHVVXUHVXFKDVZDWHUKDPPHULQJ
• The use of standard models in the following installation locations may cause
GHYLFHIDLOXUH,QWKHVHORFDWLRQVSOHDVHXVHWKH37*BBZLWKWKHRSWLRQDOIHDWXUH
of resistance to dynamic pressure or resistance to pulsation.
[Recommended locations for dynamic pressure resistant model]
:KHUHG\QDPLFSUHVVXUHLVDSSOLHGUHSHDWHGO\WRWKLVGHYLFHGXHWRDEDWFK
process
:KHUHWKHVHQVRURIWKLVGHYLFHLVH[SRVHGWRDZDWHUVKRZHUIRUFOHDQLQJD
WDQN([WRSRIDFRQLFDOWDQN
2WKHUSODFHVZKHUHG\QDPLFSUHVVXUHLVDSSOLHGWKLVGHYLFH([QHDUSLSH
EHQGV
[Recommended locations for pulsation resistant model]
:KHUHWKLVGHYLFHLVH[SRVHGGLUHFWO\WRSURFHVVSXOVDWLRQ([URWDU\SXPS
RXWOHW
• After installing the transmitter, adjust the zero point.
)RUGHWDLOVRQWKH]HURSRLQWDGMXVWPHQWPHWKRGUHIHUWRFKDSWHU³&DOLEUDWLRQ´