Do you have a question about the B. Braun Omni and is the answer not in the manual?
Ensures the OMNI remains charged and ready for use, even when not actively in therapy.
Instructions for the correct operation of the power button for machine activation and deactivation.
Details the OMNI's battery life and the impact of unplugging on therapy continuity.
Guidance on correctly attaching fluid and blood lines to the OMNI set.
Procedure for connecting pressure lines, noting machine feedback during the process.
Instructions to let the machine load the set, avoiding manual manipulation to prevent damage.
Correct identification and connection of syringe lines, particularly for Calcium Chloride and Heparin.
Information on using different fluid bags for CWHD/CVVHDF and anticoagulation types.
Importance of correctly breaking frangible pins on fluid bags to prevent alarms.
Note that scales are separate and do not require balancing.
Ensuring correct clamp status for unused and used lines before initiating automatic priming.
Steps to initiate and allow the machine to complete the automatic priming sequence.
Procedure for resolving the arterial line blocked alarm during priming.
Advice to double-check clamps and pins for flow or fluid-related priming alarms.
Parameters can be set during priming or before patient connection.
Importance of matching Calcium Chloride and Citrate concentrations to machine settings.
Guidelines on adjusting blood flow when using citrate and consulting staff for queries.
Recommendation to leave the therapy time limit blank to run for the set's lifetime.
Correct input of net fluid removal and dialysate flow values.
How to re-prime lines with saline after initial priming.
Steps for recirculating the set before commencing therapy, including clamp usage.
Starting the blood pump and adjusting flow rate based on patient stability.
Steps to resolve a low venous alarm, including checking connections and bed height.
Advice on using red/blue buttons and automatic restarts for therapy management.
Correct procedure for changing fluid bags using the 'bags' button to avoid errors.
Understanding and preventing the alarm caused by prolonged anticoagulation stops.
Procedure for safely removing air from the calcium syringe.
Causes and resolution steps for blood leak detector alarms, including overrides.
Resolving backflow of blood into the arterial pressure line through specific steps.
Diagnosing and rectifying alarms related to fluid flow, pump, or solution issues.
Reasons for OMNI clotting, emphasizing access issues and anticoagulation.
How the machine performs 24-hour self-tests and handles associated alarms.
Troubleshooting alarms related to solution chamber regulation due to pins, clamps, or condensation.
Steps for safely returning blood to the patient before ending therapy.
Procedure for unloading disposable components after therapy completion.
| Manufacturer | B. Braun |
|---|---|
| Model | Omni |
| Accuracy | ± 5% |
| Power Supply | 100-240 VAC, 50/60 Hz |
| Type | Infusion Pump |
| Infusion Modes | Continuous, Intermittent |
| Flow Rate Range | 0.1 to 999 ml/h |
| Volume Range | 0.1 to 9999 ml |
| Display | LCD |
| Alarms | Occlusion, Air-in-line, Low battery, Infusion complete |