B. Braun microspeed uni - User Manual

B. Braun microspeed uni
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Aesculap Power Systems
microspeed® uni
Instructions for use
microspeed® uni motor system
Gebrauchsanweisung
microspeed® uni-Motorensystem
Mode d’emploi
¾ voir www.aesculap-extra.net
Instrucciones de manejo
¾ ver www.aesculap-extra.net
Istruzioni per l’uso
¾ vedere www.aesculap-extra.net
Instruções de utilização
¾ ver www.aesculap-extra.net
Gebruiksaanwijzing
¾ zie www.aesculap-extra.net
Bruksanvisning
¾ se www.aesculap-extra.net
Käyttöohjeet
¾ ks. www.aesculap-extra.net
Инструкция по примению
¾ смотри www.aesculap-extra.net

Table of Contents

Questions and Answers

  • E
    Elizabeth GaySep 23, 2025
    Why is the pistol handpiece not running on my B. Braun microspeed uni?
    • H
      Hannah ArmstrongSep 23, 2025
      If the pistol handpiece isn't running, check if the protection against inadvertent switch-on is activated. If so, turn back the upper trigger.
  • J
    Jason DavisAug 23, 2025
    What causes Error 3 – Foot-Control on B. Braun Medical Equipment, and how to fix it?
    • K
      Kristin RamirezAug 23, 2025
      Error 3 on your B. Braun Medical Equipment foot control can arise from several issues: * If the pedal(s) were pressed during the automatic self-test, release them. * If using foot control GD671, ensure only one pedal is pressed unless in "Shave" mode with a shaver handpiece. * Try cleaning the plug contacts. * If none of these resolve the error, the foot control may be defective and require replacement.
  • J
    Jonathan HamptonAug 25, 2025
    What does Error 6 – Motor/Hallsensor mean on B. Braun Medical Equipment, and how to resolve it?
    • M
      Melissa StevensAug 26, 2025
      Error 6 related to the Motor/Hallsensor indicates a malfunction in the Hall system of the motor (specifically for motors GD674, GD678, GD679, GD682, and GD684). In this case, replace the motor. This error could also be caused by a defective motor cable GD672 or GD673, which would also need to be replaced.
  • C
    Christine RobertsAug 27, 2025
    What does Error 4 – Hand-Control mean on B. Braun Medical Equipment, and how can I fix it?
    • E
      Eric MoralesAug 27, 2025
      Error 4 indicates a problem with the hand control. This can happen if the lever of motor cable GD673 or the trigger of pistol handpiece GD684 was activated during the automatic test; release the lever or trigger. If that doesn't work, the motor cable GD673 or the pistol handpiece GD684 may be defective and need replacement.
  • D
    davisralphAug 27, 2025
    Why does B. Braun microspeed uni show Error 7 - Motor overheated?
    • K
      Katherine DaviesAug 28, 2025
      Error 7 indicates the motor has overheated. Allow the motor to cool down. If the error persists, the motor may need to be replaced.
  • W
    William ThomasSep 18, 2025
    What causes Error 10 – Pump on B. Braun microspeed uni?
    • A
      Amanda LewisSep 18, 2025
      Error 10 suggests a mechanical blockage of the pump. Check the position of the tubing.
  • H
    Heather FreemanSep 15, 2025
    What to do if the B. Braun Medical Equipment control unit cannot be operated and the motor does not run?
    • E
      emily05Sep 15, 2025
      If the control unit is inoperable and the motor won't run, it could be due to the internal monitoring system blocking the system. In this case, switch the control unit off and on again. If the error persists, have the control unit repaired by the manufacturer.
  • J
    joneswilliamSep 16, 2025
    Why is there no coolant flow in my B. Braun microspeed uni Medical Equipment?
    • L
      Lauren GarciaSep 17, 2025
      Lack of coolant flow can be due to several reasons: * The nozzle may be clogged, so clean it. * The tubing might be incorrectly inserted. * The coolant pump motor could be defective.
  • M
    Mrs. Diane UnderwoodSep 6, 2025
    What does Error 1 mean on B. Braun microspeed uni?
    • C
      Christine WrightSep 7, 2025
      Error 1 indicates that more than one key was pressed at the same time. Only press one key at a time.
  • H
    Hannah SimmonsSep 4, 2025
    What does Error 9 mean on B. Braun microspeed uni Medical Equipment, and how can I fix it?
    • L
      Laura JonesSep 4, 2025
      Error 9 indicates that current cannot be properly applied to the motor. Replace the motor or motor cable.

Summary

Product Overview

System Components and Intended Use

Details system components, necessary items, and intended applications of the microspeed® uni system.

Mode of Operation

Explains different modes of operation, including control unit, display interface, and motor controls.

Safe Handling

Symbols

Control Unit and Device Symbols

Illustrates symbols used on the control unit and various system components like foot controls and motors.

Display Symbols in Motor Control Fields

Explains symbols displayed in the motor control fields for status and operation indicators.

Preparation and Setup

Control Unit Mounting, Dismounting, and Power Connection

Details procedures for mounting, dismounting, and connecting the control unit to mains power.

System Operation

System Setup and Initial Connections

Covers initial system setup, connecting accessories, foot controls, tubing, and motor cables.

Control Unit Operation and Function Checks

Explains control unit power-on, automatic self-test, function checks, and basic handpiece connections.

Handpiece and Motor Operation

Describes the setup and operation of shaver and pistol handpieces, including motor activation.

Instruction Menu Settings

Information and Error Handling

Covers accessing information, error lists, and software versions within the instruction menu.

Device and Motor-Specific Settings

Explains how to adjust device-specific settings and motor type-specific parameters.

Processing and Maintenance

Processing Procedures

Details cleaning, disinfection, and sterilization methods for the control unit, foot controls, motors, and handpieces.

Maintenance Schedule

Outlines the recommended maintenance schedule and procedures for the microspeed® uni system.

Troubleshooting

Malfunctions and Error Codes

Lists common malfunctions with error numbers, their causes, and recommended detection/remedy steps.

Other Common Issues

Addresses other non-error related malfunctions and provides troubleshooting guidance.

Support, Parts, and Specs

Fuse Change, Service, and Accessories

Includes instructions for fuse changes, technical service contacts, and lists available accessories.

Spare Parts and Power Cords

Lists spare parts and power cord options for the microspeed® uni motor system.

Technical Specifications

Presents detailed technical specifications, classifications, and performance data for system components.

Disposal and Contact

Product Disposal Guidelines

Explains the proper procedures and regulations for disposing of the microspeed® uni motor system.

Distributor and Product Information

Lists contact details for product information, support, and distributor information in the US and Canada.

B. Braun microspeed uni Specifications

General IconGeneral
BrandB. Braun
Modelmicrospeed uni
CategoryMedical Equipment
LanguageEnglish