Aesculap Power Systems
microspeed® uni
Instructions for use
microspeed® uni motor system
Gebrauchsanweisung
microspeed® uni-Motorensystem
Mode d’emploi
¾ voir www.aesculap-extra.net
Instrucciones de manejo
¾ ver www.aesculap-extra.net
Istruzioni per l’uso
¾ vedere www.aesculap-extra.net
Instruções de utilização
¾ ver www.aesculap-extra.net
Gebruiksaanwijzing
¾ zie www.aesculap-extra.net
Bruksanvisning
¾ se www.aesculap-extra.net
Käyttöohjeet
¾ ks. www.aesculap-extra.net
Инструкция по примению
¾ смотри www.aesculap-extra.net

Table of Contents

Questions and Answers

  • M
    Mrs. Diane UnderwoodSep 6, 2025
    What does Error 1 mean on B. Braun microspeed uni?
    • C
      Christine WrightSep 7, 2025
      Error 1 indicates that more than one key was pressed at the same time. Only press one key at a time.
  • H
    Heather AndersonAug 22, 2025
    What to do if B. Braun microspeed uni shows Error 6 – Motor?
    • R
      Raymond GallagherAug 22, 2025
      Error 6 indicates a motor issue. This might be because no motor is connected to the control unit, so connect a motor. Alternatively, the motor itself could be defective and needs replacement, or the motor cable GD672 or GD673 may be faulty and require replacement.
  • H
    Hannah SimmonsSep 4, 2025
    What does Error 9 mean on B. Braun microspeed uni Medical Equipment, and how can I fix it?
    • L
      Laura JonesSep 4, 2025
      Error 9 indicates that current cannot be properly applied to the motor. Replace the motor or motor cable.
  • S
    Shelly MartinezSep 2, 2025
    What does Error 5 – Select only one Motor mean on B. Braun microspeed uni Medical Equipment, and how do I resolve it?
    • K
      Kathryn MillerSep 2, 2025
      Error 5 concerning motor selection indicates a speed request from the foot control while two motors are released for operation (motor release switch on motor cable GD672 is in the "ON" position). Release the required motor using the motor release switch (position "ON") and block the motor that is not required using the motor release switch (position "OFF"). The motor cable GD672 may also be defective and require replacement.
  • R
    Richard MartinezSep 1, 2025
    What causes Error 8 – Power Fail on B. Braun microspeed uni Medical Equipment?
    • E
      ethanmurraySep 1, 2025
      Error 8 signals a power supply error. Verify that the mains voltage is within the defined range.
  • D
    dwaynevargasAug 30, 2025
    What to do if B. Braun Medical Equipment displays Error 6 – Motor blocked?
    • S
      simmonsbillyAug 30, 2025
      If Error 6 indicates a blocked motor, stop motor actuation and release the blockage. If the error continues, the motor may be defective and require replacement.
  • L
    laura98Aug 29, 2025
    What does Error 5 – Select Motor mean on B. Braun microspeed uni Medical Equipment and how to fix it?
    • G
      Gregory CalderonAug 29, 2025
      Error 5 regarding motor selection means that there is a speed request from the foot control while the motor is locked by the motor release switch on motor cable GD672 (position "OFF"). Set the motor release switch on the motor cable to the "ON" position. Alternatively, the motor cable GD672 may be defective and require replacement.
  • D
    davisralphAug 27, 2025
    Why does B. Braun microspeed uni show Error 7 - Motor overheated?
    • K
      Katherine DaviesAug 28, 2025
      Error 7 indicates the motor has overheated. Allow the motor to cool down. If the error persists, the motor may need to be replaced.
  • C
    Christine RobertsAug 27, 2025
    What does Error 4 – Hand-Control mean on B. Braun Medical Equipment, and how can I fix it?
    • E
      Eric MoralesAug 27, 2025
      Error 4 indicates a problem with the hand control. This can happen if the lever of motor cable GD673 or the trigger of pistol handpiece GD684 was activated during the automatic test; release the lever or trigger. If that doesn't work, the motor cable GD673 or the pistol handpiece GD684 may be defective and need replacement.
  • J
    Jonathan HamptonAug 25, 2025
    What does Error 6 – Motor/Hallsensor mean on B. Braun Medical Equipment, and how to resolve it?
    • M
      Melissa StevensAug 26, 2025
      Error 6 related to the Motor/Hallsensor indicates a malfunction in the Hall system of the motor (specifically for motors GD674, GD678, GD679, GD682, and GD684). In this case, replace the motor. This error could also be caused by a defective motor cable GD672 or GD673, which would also need to be replaced.

Summary

Product Overview

System Components and Intended Use

Details system components, necessary items, and intended applications of the microspeed® uni system.

Mode of Operation

Explains different modes of operation, including control unit, display interface, and motor controls.

Safe Handling

Symbols

Control Unit and Device Symbols

Illustrates symbols used on the control unit and various system components like foot controls and motors.

Display Symbols in Motor Control Fields

Explains symbols displayed in the motor control fields for status and operation indicators.

Preparation and Setup

Control Unit Mounting, Dismounting, and Power Connection

Details procedures for mounting, dismounting, and connecting the control unit to mains power.

System Operation

System Setup and Initial Connections

Covers initial system setup, connecting accessories, foot controls, tubing, and motor cables.

Control Unit Operation and Function Checks

Explains control unit power-on, automatic self-test, function checks, and basic handpiece connections.

Handpiece and Motor Operation

Describes the setup and operation of shaver and pistol handpieces, including motor activation.

Instruction Menu Settings

Information and Error Handling

Covers accessing information, error lists, and software versions within the instruction menu.

Device and Motor-Specific Settings

Explains how to adjust device-specific settings and motor type-specific parameters.

Processing and Maintenance

Processing Procedures

Details cleaning, disinfection, and sterilization methods for the control unit, foot controls, motors, and handpieces.

Maintenance Schedule

Outlines the recommended maintenance schedule and procedures for the microspeed® uni system.

Troubleshooting

Malfunctions and Error Codes

Lists common malfunctions with error numbers, their causes, and recommended detection/remedy steps.

Other Common Issues

Addresses other non-error related malfunctions and provides troubleshooting guidance.

Support, Parts, and Specs

Fuse Change, Service, and Accessories

Includes instructions for fuse changes, technical service contacts, and lists available accessories.

Spare Parts and Power Cords

Lists spare parts and power cord options for the microspeed® uni motor system.

Technical Specifications

Presents detailed technical specifications, classifications, and performance data for system components.

Disposal and Contact

Product Disposal Guidelines

Explains the proper procedures and regulations for disposing of the microspeed® uni motor system.

Distributor and Product Information

Lists contact details for product information, support, and distributor information in the US and Canada.

B. Braun microspeed uni Specifications

General IconGeneral
BrandB. Braun
Modelmicrospeed uni
CategoryMedical Equipment
LanguageEnglish