What causes Error 8 – Power Fail on B. Braun microspeed uni Medical Equipment?
- EethanmurraySep 1, 2025
Error 8 signals a power supply error. Verify that the mains voltage is within the defined range.
What causes Error 8 – Power Fail on B. Braun microspeed uni Medical Equipment?
Error 8 signals a power supply error. Verify that the mains voltage is within the defined range.
What to do if B. Braun Medical Equipment displays Error 6 – Motor blocked?
If Error 6 indicates a blocked motor, stop motor actuation and release the blockage. If the error continues, the motor may be defective and require replacement.
What does Error 5 – Select Motor mean on B. Braun microspeed uni Medical Equipment and how to fix it?
Error 5 regarding motor selection means that there is a speed request from the foot control while the motor is locked by the motor release switch on motor cable GD672 (position "OFF"). Set the motor release switch on the motor cable to the "ON" position. Alternatively, the motor cable GD672 may be defective and require replacement.
Why does B. Braun microspeed uni show Error 7 - Motor overheated?
Error 7 indicates the motor has overheated. Allow the motor to cool down. If the error persists, the motor may need to be replaced.
What does Error 4 – Hand-Control mean on B. Braun Medical Equipment, and how can I fix it?
Error 4 indicates a problem with the hand control. This can happen if the lever of motor cable GD673 or the trigger of pistol handpiece GD684 was activated during the automatic test; release the lever or trigger. If that doesn't work, the motor cable GD673 or the pistol handpiece GD684 may be defective and need replacement.
What does Error 6 – Motor/Hallsensor mean on B. Braun Medical Equipment, and how to resolve it?
Error 6 related to the Motor/Hallsensor indicates a malfunction in the Hall system of the motor (specifically for motors GD674, GD678, GD679, GD682, and GD684). In this case, replace the motor. This error could also be caused by a defective motor cable GD672 or GD673, which would also need to be replaced.
What causes Error 3 – Foot-Control on B. Braun Medical Equipment, and how to fix it?
Error 3 on your B. Braun Medical Equipment foot control can arise from several issues: * If the pedal(s) were pressed during the automatic self-test, release them. * If using foot control GD671, ensure only one pedal is pressed unless in "Shave" mode with a shaver handpiece. * Try cleaning the plug contacts. * If none of these resolve the error, the foot control may be defective and require replacement.
Brand | B. Braun |
---|---|
Model | microspeed uni |
Category | Medical Equipment |
Language | English |
Details system components, necessary items, and intended applications of the microspeed® uni system.
Explains different modes of operation, including control unit, display interface, and motor controls.
Illustrates symbols used on the control unit and various system components like foot controls and motors.
Explains symbols displayed in the motor control fields for status and operation indicators.
Details procedures for mounting, dismounting, and connecting the control unit to mains power.
Covers initial system setup, connecting accessories, foot controls, tubing, and motor cables.
Explains control unit power-on, automatic self-test, function checks, and basic handpiece connections.
Describes the setup and operation of shaver and pistol handpieces, including motor activation.
Covers accessing information, error lists, and software versions within the instruction menu.
Explains how to adjust device-specific settings and motor type-specific parameters.
Details cleaning, disinfection, and sterilization methods for the control unit, foot controls, motors, and handpieces.
Outlines the recommended maintenance schedule and procedures for the microspeed® uni system.
Lists common malfunctions with error numbers, their causes, and recommended detection/remedy steps.
Addresses other non-error related malfunctions and provides troubleshooting guidance.
Includes instructions for fuse changes, technical service contacts, and lists available accessories.
Lists spare parts and power cord options for the microspeed® uni motor system.
Presents detailed technical specifications, classifications, and performance data for system components.
Explains the proper procedures and regulations for disposing of the microspeed® uni motor system.
Lists contact details for product information, support, and distributor information in the US and Canada.