46
microspeed® uni motor system
Aesculap Power Systems
7. Troubleshooting list
7.1 Malfunctions with error numbers
As soon as an error status is detected, an error window will pop up on the
display and an acoustic warning will be issued.
In the event a malfunction cannot be remedied as described below:
¾ Have component(s) repaired by the manufacturer.
Malfunction Message displayed Cause Detection/Remedy
Error 1 Error 1 Key actuation (softkey actuation):
More than one key was pressed at the same
time
Only press one key
Error 2 Error 2 Key (softkey) held down for too long:
Control key pressed for longer than 20 s
Do not press any key for longer than 20 s
Error 3 Error 3 – Foot-Control Pedal(s) of the foot control pressed during
automatic self-test
Release pedal(s)
Both pedals of foot control GD671 pressed Press one pedal only
Pressing both pedals is only allowed in
operating mode "Shave" in conjunction
with the shaver handpiece
Plug contacts of foot control unclean Clean the plug contacts
Foot control defective Replace foot control
Error 4 Error 4 – Hand-Control Lever of motor cable GD673 or trigger of
pistol handpiece GD684 actuated during
automatic test
Release lever or trigger, respectively
Motor cable GD673 defective Replace motor cable
Pistol handpiece GD684 defective Replace pistol handpiece
Error 5 Error 5 – Select Motor Speed request from foot control while the
motor is locked by the motor release switch
on motor cable GD672 (position "OFF")
Set motor release switch on motor cable to
position "ON"
Motor cable GD672 defective Replace motor cable
Error 5 – Select only one Motor Speed request from foot control while two
motors are released for operation the motor
release switch on motor cable GD672
(position "ON")
Release the required motor through motor
release switch (position "ON")
Block the motor that is not required
through motor release switch (position
"OFF")
Motor cable GD672 defective Replace motor cable
Error 5 – Connect only one Shaver Speed request from foot control while two
shaver handpieces GD682 are connected to
the control unit
Connect only one shaver handpiece GD682
to the control unit