Vista
basic
Infusion Pump
Quick
Reference
Rx only
Vista™ and Working Together for Excellence in Safety
are a trademarks of B. Braun Medical Inc.
©2002, B. Braun Medical Inc.
950853-Rev B 06/02
Alarm Troubleshooting
Air In Line! Check for air in air-in-line
detector.
Check that tubing is properly
inserted in air-in-line detector.
Remove air from line.
Battery Plug pump into wall outlet.
Too Low!
Door Open! Close door.
Dwnstream Open clamps.
Occlusion!
Check for kinked tubing.
Check IV site.
Check occlusion pressure
setting.
Pump Pump is stopped.
Stopped!
Either start infusion or press
to extend hold time
3 minutes.
Silence alarms by pressing
.
Refer to Instructions for Use for complete
operating instructions.
B. Braun Medical Inc.
824 Twelfth Avenue
Bethlehem, PA 18018
1-800-227-2862
www.bbraunusa.com
Program Mode
1. Press OPTS until
PROGRAM
MODE
is
displayed
2. Press ON
.
3. Enter the rate for Step 1 and press OK
.
4. Enter the volume for Step 1 and press OK
.
5. Repeat Steps 3 and 4 until all steps are
complete and press OK
.
6. Review the Total VTBD and press OK
.
7. Press START
.
Ramp Taper Mode
1. Press OPTS until
RAMP
TAPER
MODE
is
displayed
2. Press ON
.
3. Enter the total bag volume to be delivered and
press OK
.
4. Enter the Ramp Up time in hours and minutes
and press OK
.
5. Enter the Taper Down time in hours and minutes
and press OK
.
6. Enter the Total Time for the infusion in hours
and minutes and press OK
.
7. Review the calculated Max Rate and
press OK
.
8. Press START
.

Other manuals for B. Braun Vista basic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the B. Braun Vista basic and is the answer not in the manual?

B. Braun Vista basic Specifications

General IconGeneral
TypeInfusion Pump
ManufacturerB. Braun
ModelVista basic
Flow Rate Range0.1 to 999 mL/h
Volume Range1 to 9999 mL
Accuracy±5%
DisplayLCD
CompatibilityStandard infusion sets
Infusion ModesContinuous
Power Supply100 – 240 V AC, 50/60 Hz
AlarmsAir-in-line, Occlusion, Low Battery
Weight4.4 lbs (2 kg)