ューー
Software
Control
Microprocessor
68hc11
FFxx
is
displayedin
the
7
segment
display
with
flashing
dots.
FFxx
is
the
error
code.
7
Segment
Display
Description
ΕΕΟΙ
dummy
for
test
FFO2
battery
not
present
/
missing
battery
current
FFO3
defective
RAM
memory
FF04
defective
program
memory
-
ROM
test
error
1
FF05
defective
program
memory
-
ROM
test
error
2
FF06
calibration
data
error
from
EEPROM
FF07
pump
head
cycle
not
plausible
FFO8
failure
/
inaccuracy
of
system
clock
FFO9
failure
100msec
system
clock
FF10
reset
during
active
operation
FF12
no
dynamic
pressure
sensor
signal
(EDB)
FF14
defective
temperature
sensor
FF16
defective
membrane
keypad
FF17
defective
program
memory
tube
parameters
Table
2
-
2