6
Reporting Procedure
50
DHV0078 b2 EN
© b2 electronic GmbH
Steps R1-R13 list the report information that the HVA asks the user to enter
when the “Extended” reporting mode is active.
Step
Procedure
(Reporting)
Example
R1:
DUT
Specify the type of DUT. Select one of the following
options:
Cable
Motor
Generator
Transformer
Switchgear
Other
Vacuum bottle
R2:
Voltage rating
Specify the voltage rating of the DUT.
This is a characteristic of the DUT and does NOT
refer to the test voltage!
Rotate navigation knob to increase or decrease
voltage rating value.
R3:
Insulation
(only applicable for
cables)
Specify one of the following cable insulation types:
XLPE
PILC
EPR
PE
PVC
HYBRID
(combination of types)
OTH. (other)
R4:
Report title
Set report name
User defined entry, typically the cable number or
ID for cable testing
R5:
Phase
(Extended only)
Specify circuit phase
User can define up to 3 phases if required
R6:
Company
(Extended only)
Specify company name
Start a New Test
What is the
Device Under Test?
Cable
MAIN PREV NEXT
Start a New Test
What is the
Voltage Rating of DUT?
1.0kV
MAIN PREV NEXT
DUT: Cable
Start a New Test
What type of
Insulation?
XLPE
MAIN PREV NEXT
DUT: 1.0kV Cable
Start New / Report Details
Report Title
Circuit / Line ID
circuitALPHA
MAIN PREV NEXT
DUT: 1.0kV Cable XLPE
Start New / Report Details
Phase?
phaseRED
MAIN PREV NEXT
DUT: 1.0kV Cable XLPE
Start New / Report Details
Company Name?
company HV
MAIN PREV NEXT
DUT: 1.0kV Cable XLPE