5
Test Procedure
46
DHV0078 b2 EN
© b2 electronic GmbH
5.4 Interrupting a Test
Once a test has started, it can be interrupted at any time. It is recommended to
select the appropriate method corresponding to the situation.
Situation
Procedure
Example
Routine STOP
(No emergency)
When a test is in progress, STOPon the display screen
is highlighted.
To interrupt the test, push in/click the navigation knob .
HVA software deactivates HV
Test stops
Alternative
When a test is in progress, press the HV switch to
deactivate high voltage.
HVA hardware deactivates HV
Test stops
Emergency stop
In an emergency situation, press the emergency OFF
to shut down the system.
HVA hardware deactivates HV
Test stops
After test interruption, a message is displayed indicating that the test has been
terminated by the user:
Manual Mode Sine 0.1Hz
STOP T: 00:58 / 1min
1.0
kV
rms
0.1kV
0.9µA
0.4nF
13GΩ
Manual Mode Finished
Manual Mode Test Seq.
interrupted by the user at
step 1 after 58s
Sine 0.1 Hz
Test Voltage: 1.0 kVrms
OK