Average calculation will be time and date coded. Pressure
and Temperature records will also be shown but can not be
selected for the Average calculations.
4. Use the ▲▼ buttons to highlight any three consecutive combustion
records, then press ENT to display the calculated average for all data
stored in the records.
NOTE: If the highlighted records contain a Temperature or
Pressure record, the calculation will not be performed.
5. Press the print key (F1) to print a copy of all three records and their
average, or ESC to return to the Print Average Screen.
Figure 5-22. Print Average
5.14. Downloading Stored Data to a Computer
The combustion test data that was stored in either the analyzer’s Memory
Directory or Logging Directory (see Memory Functions section on page 58), can
be downloaded to a computer using the Fyrite
®
User Software and USB cable
that are supplied with the analyzer.
The downloaded data is stored on the computer’s hard drive – or removable
media of the operator’s choosing – as a comma-separated-value ASCII text file
with a CSV extension. This type of file can be opened by most spreadsheet
programs for analysis.
For information on the installation of the Fyrite
®
User Software, refer to the
installation document (provided in .PDF format) on the software disk. The Fyrite
®
User Software installation manual may also be downloaded from the PCA
®
3
page of the Bacharach website (
www.MyBacharach.com). Refer to manual
0024-9465. Minimum computer requirements for installation and use of the
Fyrite
®
User Software are shown below.