Bacharach PCA3 - 6.11. COHIGH Sensor Calibration

Bacharach PCA3
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PCA
3 Manual
76 0024-9473 Rev 3
6.8. SO
2
Sensor Calibration
This procedure spans the optional sulfur dioxide sensor to a known gas
level.
6.8.1. Material Required
Calibration Kit ( P/N 0024-7059)
Gas Cylinder (100 ppm SO
2
in Nitrogen, P/N 0024-1158)
6.8.2. Procedure
1. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 66).
2. Use the ▲▼ buttons to highlight SO
2
, and then press ENT to display
the CALIBRATE SO
2
screen.
NOTE: “Measured” is the current SO
2
reading, while
“Applied” is a known SO
2
level that will be applied for
calibration purposes.
3. Attach a 100 ppm SO
2
cylinder to the regulator of the calibration fixture
shown in Figure 6-12 on page 80.
4. Use the ▲▼ and
◄►
buttons to enter an applied value that exactly
equals the concentration stamped on the SO
2
cylinder.
NOTE: The calibration range is from 9 to 1,800ppm. An
attempt to calibrate outside this range will cause the
message “Bad Calibration Wrong CAL Entry” to appear in
the following step.
Figure 6-9. SO
2
Sensor Calibration
5. Adjust regulator for a flow meter indication of approximately 2  SCFH.
Wait until the measured reading stabilizes (approximately 3  minutes),
and then press ENT to calibrate the SO
2
Measured value to that of the
applied value. The message “Good Calibration” should briefly appear
followed by the CALIBRATION LIST screen being re-displayed.