Document: 333D012G
5-18-2007
- 50 -
IMPORTANT
Before replacing a UV germicidal lamp it might be necessary to check with an industrial
hygienist, safety officer or other qualified person in the lab to make sure that there is nothing
stored in the cabinet that could potentially be a hazard to you or could be contaminated by
room air.
1) Raise the viewscreen to its maximum position. The audible alarm should activate and the indicator on
the light canopy will flash. To mute this audible alarm while the UV lamp is being replaced, press the
alarm reset button located on the light canopy. After five minutes the alarm will sound again and you
may press the reset button if more time is needed for lamp replacement.
2) Using both hands, grasp the UV lamp carefully near each end and move it approximately ½” [13mm] to
the right side of the cabinet until the UV lamp is released from the lamp socket on the left side.
Carefully remove lamp from cabinet and dispose of it properly.
3) Install new lamp by following step 2 in reverse order. Making sure that the lamp is held securely in
place.
4) Lower the viewscreen to full closed and turn on the UV light momentarily to verify that it comes on.
Immediately turn the UV off again.
5) Raise the viewscreen to the proper sash opening height. This will cancel the audible alarm and flashing
indicator light.