Document: 333D012G
5-18-2007
- 51 -
Installation of Exhaust Transition Systems for SterilGARD
®
III Cabinets
Tools required:
5/16 - Hex wrench and nut driver
11/32 - Hex wrench and nut driver
3/8 - Hex wrench and nut driver
7/16 - Hex wrench and nut driver
11/32 - Combination wrench
Unpacking the Transition:
1) Remove contents from the shipping packaging and examine for damage. (Be careful not to damage
the gasket on the bottom flange of the exhaust transition)
2) Check the quantity of the components to the following packing lists and illustrations:
Item # Description
Qty
SG403A
Qty
SG 603A
1 Exhaust transition 1 1
2 Exhaust filter access cover (closed face) 1 1
3 ¼”-20 x ¾” Washer head bolt 4 6
4 ¼” Seal washer 6 8
5* ¼”-20 x 1” Hex head bolt 4 4
6* ¼” Flat washer 6 6
7* ¼” Lock washer 6 6
8* ¼”-20 Double ended transition mounting bolt 2 2
9* ¼”- 20 Hex nut 2 2
10* Exhaust transition mounting clip 2 2
11* Air tight damper assembly (ATD) 1 1
12* Short exhaust collar (Optional) 1 1
13* #8-32 Cap nut 8 8
14* #8 Flat washer 8 8
15* #8 Lock washer 8 8
16 Exhaust transition gasket, ¼” x ½” 8.3 ft 10.3 ft
17 Access cover gasket, ¼” x ½” 6.2 ft 8.2 ft
18 Exhaust collar gasket, 1/8” 1 1
Packing List for Hard Exhaust Connection (HEC)
* = Items shipped loose in the packaging.