Towel guard
The towel guards are located under the work surface at the bottom rear and sides of the return-air plenums. Acting
as a protective screen, integral to the interior walls, they help prevent wipes and other paper materials from being
drawn into the blower system. They need to be kept clean at all times.
All-metal plenums
The plenums are constructed entirely of carbon and stainless steel in order to provide strength, durability, air-
tightness and resistance to deterioration.
Removable recessed stainless steel work surface
The work surface is constructed of corrosion resistant 16-gauge type 304 stainless steel, with a satin finish that
diminishes light reflection. It is recessed to retain spills and can be removed along with its supports to gain access to
the drain pan.
Drain pan
The drain pan is designed with 7/16” [11.1mm] radius in all four bottom corners to facilitate cleaning and
disinfection. Drainage is provided by a 1/2” [12.7mm] diameter stainless steel ball valve located at the bottom of the
drain pan.
Viewscreen
The cabinet’s 1/4” [6.35mm] safety plate glass sliding viewscreen may be opened to 19 3/8” [492.1mm] for placing of
large items in the work area, and maybe fully closed for system shutdown or UV light operation.
Work area lighting
The work area is illuminated by two external fluorescent lamps which provide 100 foot-candles of light at the
work surface.
Electronic ballast
The SterilGARD
®
III features solid-state electronic ballasts for the fluorescent and UV (optional) lights. These
ballasts increase reliability, efficiency, and service life with lower heat output.
Adjustable cabinet height
The cabinet has adjustable legs and leg levelers. The legs provide 6” [152.4mm] of height adjustment and the leg
leveler provides an additional 2.5” [63.5mm] of height adjustment.
Sponge armrest pad
The armrest pad is made out of EPDM sponge material and is resistant to most chemicals and UV exposure. It is
held in place with a low tack adhesive so the pad can be easily removed for cleaning. It also can be autoclaved.
UniPressure plenum
A telescoping all steel positive pressure plenum provides a more even clamping force on the HEPA filter frames
and helps deliver the down flow air more uniformly. The plenum can be easily telescoped for quick filter changing.