Monitor display
Error
No.
Signifi cance Cause / Measure Remedial action
Error !!! 003
please
contact service
please
contact service
003 No power output
Device fault
Repairs only
through Customer
Service
- Send in homogenizer for repair
since fault is severe
Error !!! 011
please
check the probe
002
Frequency setting not
possible
Ultrasonic
transducer or probe
faulty. Replace
parts and check
function (conduct
a “probe check” or
“scan frequency”)
If no success
Repairs only
through Customer
Service.
- Check the connection to the
probe.
Is the cable properly plugged
in and affi xed? Is the probe
correctly screwed on? Check
for erosion on the probe.
- Once error rectifi ed, conduct
the “probe check” or “scan
frequency” function (see
chapter 6.2).
011
No response signal
from the ultrasonic
transducer
012
No resonance
frequency found
Error !!! 014
max. temperature
exceeded
014
Permissible internal
operating temperature
exceeded. Ultrasonic
homogenizer turns off.
Allow the ultrasonic
homogenizer to
cool down.
- Wait until the HF generator
has suffi ciently cooled. The
operating temperature must be
< 50 °C.