8 SEGMENT & 4X16 SEGMENT PROGRAMMABLE MODELS W/OTP
dwEll
To maintain a constant temperature for a specified time:
1. Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2. Press the SCROLL button until display reads,
“tYPE.”
3. Press the UP or DOWN button until display
reads, “dwEll.”
4. Press the SCROLL button until display reads,
“dur.”
5. Press the UP or DOWN button to select a time in
dwell units (seconds, minutes or hours; set in the
“dwL.U” parameter).
StEP
To climb instantaneously from the current temperature to a
specified temperature.
1. Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2. Press the SCROLL button until display reads,
tYPE.”
3. Press the UP or DOWN button until the display
reads, “StEP.”
4. Press the SCROLL button until display reads,
“tGt.”
5. Press the UP or DOWN button to set a target set-
point.
cALL (Running Multiple Programs; 4x16 Segment
Programmable Models Only)
To call a program as a subroutine:
If you want to run multiple programs, you can program the
controller to “call” or link one program to another. This
makes it possible to run one program at any time during
another program and also return to the original program if
desired.
1. Press the PAGE button until you reach the pro-
gram list (ProG LiSt).

Table of Contents