Operation
36
Operating Instructions Radiation sensors R3XX
Ambient temperature alarm
As soon as the inside temperature of the radiation sensor exceeds the
defined value, the temperature display will start flashing and the analogue
output will switch to the programmed state (see fault current).
20.0 (corresponds to 68 °F)
70.0 or 125.0
(corresponds to 158.0 or 257.0°F)
Sensor-related: 65.0 or 125.0 °C
(corresponds to 149.0 or 257.0 °F)
Fault current
This is where you define what current is to be output via the analogue output
in the event of a fault.
The current is set in mA.
Unit of measurement
You can choose °C or °F as the unit of measurement for the temperature
display.
HART
®
address
You can operate up to 15 transmitters in parallel (multidrop mode). Each
sensor (transmitter) then requires an individual address between 1 and 15.
The address must be set to 0 if you want to operate the transmitter in
standalone operation (point-to-point operation).