HART® protocol
62
Operating Instructions Radiation sensors R3XX
34 Write damping value for the PV
float
average value 0…999.9 s
float
average value 0…999.9 s
35 Write range values for the PV
byte
range unit code
float
upper range value (MBE)
float
lower range value (MBA)
byte
range unit code
float
upper range value (MBE)
float
lower range value (MBA)
38 Reset "config changed" flag
40 Enter/exit fixed current mode loop
test (fix the analogue current at
specified value
float
current
(0 = exit fixed current mode)
float
Emissivity 0.100…1.000
float
Reflectivity 0.100…1.000
float
Transmissivity 0.100…1.000
136 Read max/min hold time
float
Max hold time [s]
float
Min hold time [s]
137 Write max/min hold time
float
Max hold time 0.0 … 999.9 s
float
Min hold time 0.0 … 999.9 s
float
Max hold time [s]
float
Min hold time [s]
144 Special command (read only)
see description
float
Value 1
float
Value 2
float
Value 1
float
Value 2
145 Special command
see description
float
Value 1
float
Value 2
float
Value 1
float
Value 2
146 Read factory settings and write
them to EEPROM
int
Password
Password [32] >only RAM
Password [34] >EEPROM
int
Success message (0 = error ; 1 =
ok)