HART® protocol
64
Operating Instructions Radiation sensors R3XX
Description of command 145
Command
Data in the instruction
float
float
Data in the reply
float
float
145
0
Value
Describe service register no. 0
0
Content of service register
145
1
Value [0-65536]
Describe service register no. 1
1
Content of service register
145
2
Value [0-65536]
Describe service register no. 2
2
Content of service register
145
3
Value [0-65536]
Describe service register no. 3
3
Content of service register
145
4
Value [0-65536]
Describe service register no. 4
4
Content of service register
145
5
Value [0-65536]
Describe service register no. 5
5
Content of service register
145
6
Value [0-65536]
Describe service register no. 6
6
Content of service register
145
7
Value [0-65536]
Describe service register no. 7
7
Content of service register
145
8
Value [0-65536]
Describe service register no. 8
8
Content of service register
145
9
Value [0-65536]
Describe service register no. 9
9
Content of service register
Upon error:
999990 = checksum incorrect
145
10
Value [0-65536]
Check sensor data and write to EEPROM
Password [34]
10
Content of service register
Upon error:
999990 = checksum incorrect
999991 = errors when writing to
EEPROM
145
146
Value [0…65536]
Read factory settings and write them to EEPROM
Password [32] >only RAM
Password [34] >EEPROM
146
Return value
0 = error
1 = ok