Page 47 of 82 OM0408/49
The sensor is calibrated to display volume magnetic susceptibility (c
vol
) for a 22mm hole.
For larger hole diameters, the displayed value will be lower than true c
vol
. An approximate value
of c
vol
can be obtained by multiplying the displayed reading by the scale factor in Table 4.
Table 4. Scaling factors for c
vol
.
Hole diameter (mm) Scale factor * for c
vol
23.0 1.000
24.0 1.418
25.4 1.739
* Average value determined by tests on 20 samples.
Figure 17. MS2K high stability surface scanning sensor.
Key to Figure 17
1. Sensing area
2. Pistol grip
3. TNC axial cable connector