BARTINGTON INSTRUMENTS
Page 64 of 82 OM0408/49
Figure 28. Water cooling system for susceptibility/temperature system
Key to Figure 28
1. Reservoir 6. Power supply to water pump
2. Power supply to liquid flow sensor 7. PVC tubing
3. Direction of flow from upper nipple on
MS2W30
8. Direction of flow to lower nipple
on MS2W30
4. Liquid flow sensor 9. Water pump
5. To reservoir 10. Water in
Note: The flow sensor must be connected to the pipe coming from the MS2W30 outlet pipe.
Note: The water pump may need to be primed when the system is initially set up. Before
connecting to the MS2W lower 6mm nipple, immerse the pump in the reservoir, and suck
water from the reservoir and through the pump into the inlet flow pipe.
Caution: It is the user’s responsibility to ensure continuity of water supply during operation
of the furnace. Failure in the water supply at elevated furnace temperatures will result in
damage to the MS2W probe.
WARNING: Do not switch on the mains electricity supply until:
1. All connections are made.
2. Power supply controls are set to RESET and HOLD and the 10-turn dial is set to zero.