BARTINGTON INSTRUMENTS
Page 50 of 82 OM0408/49
14.3.2. Taking Measurements
The sensor should be allowed to acclimatise before taking measurements. Prior to use, always
measure the calibration sample to ensure that the sensor is working correctly.
The sensor is zeroed to ‘air’ by raising the sensor about 20mm above the measurement surface
and pressing the 'Z' button. (See figure 18 for the MS2K depth response). When the zero period
is completed, the sensor can be gently placed against the surface to be measured and the 'M'
button pressed.
Move on to the next point and take a second measurement. Check the baseline drift by
performing a measurement to air.
Note: If the drift level in unacceptable then further time should be allowed for temperature
equilibration.
Note: Drift compensation is automatically accomplished using Multisus software.
14.3.3. Maintenance
Ensure that the connector is kept clean at all times, and inspect the connecting cable
occasionally for signs of wear.