Invasive Blood Pressure Warnings, Cautions, and Notes
WARNING! Avoid conductive connection with any metal parts.
NOTE! The IBP ZERO NEEDED message is displayed when the monitor is turned on, when the site label
is changed, or when the transducer is connected to the monitor (even if the same transducer is
disconnected then reconnected to the monitor).
NOTE! Use only invasive pressure transducers and interface cables specically intended for use with this
device and its side panel connectors.
NOTE! The IBP1 and IBP2 waveforms must be visible and adjacent in order to be overlaid.
NOTE! When IBP1 and IBP2 are overlaid, the scale settings for IBP1 and IBP2 are changed to the higher of
the two scales.
NOTE! Dual channel simulators may aect verication of IBP operation. Use only single channel
simulators.
NOTE! All specied transducers were tested for immunity to radiated radio frequency electromagnetic
elds at a level of 3V/m in accordance with IEC 60601-2-34:2000. Additional validation at levels
of 10V/m was performed using the pvb Critical Care GmbH xtrans
®
and Edwards Lifesciences
TruWave transducers.
Temperature Warnings, Cautions, and Notes
NOTE! Use only temperature sensors and interface cables specically intended for use with this device.