Chapter 2: Routine Maintenance
2-6 Advisor
®
Service Manual
Critical Failure Alarm (CFA) Verication
This alarm should be tested annually:
Turn the rotary knob on the monitor to move the cursor. On the main menu at the bottom of the display, 1.
highlight SETUP and push the knob to select.
Highlight SERVICE MENU and push the knob to access the service menu.2.
The service menu is password protected. To enter your password:•
Turn the rotary knob to access the password box and push the knob to select the rst character eld.a.
Turn the rotary knob to highlight the desired character and push the knob to select. The factory-b.
installed password is ADVISOR.
Push the knob to select the next character eld.c.
Turn the rotary knob to highlight the desired character and push the knob to select. Repeat steps b d.
and c until each character in the password is selected.
Highlight ENTER and push the knob to select.e.
If you correctly selected the password, the service menu will appear on the lower left corner of the •
display.
If you did not select the correct password, ***Invalid Password; please re-enter*** will be displayed in the •
message line. Try to access the service menu again.
On the service menu, highlight TEST CFA ALARM and push the knob to select.3.
Highlight YES and push the knob to select.4.
Verify that the alarm sounds.5.
To silence the critical failure alarm (CFA), press the alarm silence key (6.
B
) or highlight NO under TEST CFA
ALARM and push the knob to select.
If the alarm does not sound, contact your authorized repair center for help.