BD Phoenix™ M50 System Quick Reference Guide
To Review Panels That Need Attention
From the Status screen, select the Needs Attention tab.
Test Aborted – one of several incubator System Alerts occurred; or an ongoing panel that was in the instrument has not tested for more than one hour; or the instrument was
turned off for more than one hour; or the instrument door has been opened for more than one hour; or system software did not execute testing algorithms for more than one
hour; or a panel was moved to a different tier/instrument or Media type not specified for yeast ID panel.
Can Not Identify Barcode – internal barcode scanner could not read a panel sequence number in the same station where the instrument could determine a panel present
condition (possible scratched or missing label); or invalid panel sequence number was scanned; or an unknown panel type was scanned.
Can Not Read Panel Wells – internal barcode scanner has read a sequence number in a station but the instrument cannot determine that a panel is present in that same
station.
Panel Lot Expired – panel was logged in (or has a test start date) with a panel lot number that has already expired.
Invalid AST Results – at least one MIC field cannot be interpreted.
Panel Missing – internal barcode scanner reads a sequence number on a panel with an ongoing status during an inventory scan, but either the panel was removed before it
completed its test protocol and hasn’t obtained a result, or an internal scanner failure occurred and the sequence number can no longer be read.
No Growth on Panel – instrument did not detect growth in Growth Control well of panel; excludes QC panels and ID panels.
Review QC Results – the status of a completed QC panel is REVIEW. QC panel which yielded an incorrect ID or incorrect AST result for at least one antibiotic, or has no
growth in growth control well.
Panel Lot Undefined – Panel being logged in or placed in the instrument is from an undefined lot.
Missing Accession Number – panel does not have any accession/isolate information; excludes QC panels.
Missing Organism ID – panel has no organism ID, or has unresolved tie IDs, and has no related panel with an organism ID (which includes related panels that have unresolved
tie IDs or organism conflicts); excludes QC panels.
Can Not Determine Organism ID – panel has an Instrument Organism ID of No Identification or a related panel has No Identification for its final ID.
Invalid Organism ID – panel receives download of an organism ID not in the BD Phoenix™ database.
Organism ID Conflict – panel completes testing and has at least one related un-finalized panel that contains a different Final ID; excludes QC panels. Note that this code can
be erroneously set as a result of conflicting system activities, such as when a panel completes testing and begins auto associating ID results as the system is modifying all the
related panels with a user selected ID. If this occurs prior to the completion of the manually triggered modifications, the system can set this Needs Attention reason because
IDs were in conflict due to two independent processes occurring. This can be determined by reviewing all panel ID results.
BDXpert Rule Flagged – panel triggered at least one BDXpert rule and the rule is manually enabled in Configuration.
Pending Too Long – panel has not been scanned internally by the instrument (during an inventory scan) within 30 minutes of being entered in Panel Login.