BD Phoenix™ M50 System Quick Reference Guide
W803 Panel Lot Range Conflict Attempt was made to save panel lot but one or both panel sequence numbers conflicted with an existing
panel lot definition. A panel sequence number can only belong to one defined/saved panel lot.
W806 Panel Lot Expired A panel lot number barcode is scanned containing an expired date. Enter an unexpired panel log
number barcode.
W900 Commission Duplicate Serial
Number
There is an instrument with the same serial number already commissioned. Enter a unique serial
number.
W901 Commission Duplicate
Instrument Identifier
There is an instrument with the same designation already commissioned. Check for instrument
designation duplication.
W904 Commission Invalid Serial
Number
The serial number must be a minimum of six characters. The serial number must be six characters.
W905 Decommission of Connected
Instrument
An attempt is being made to decommission an instrument. Disconnect the instrument prior to
decommissioning.
Code Possible Cause
Corrective Action
(If the recommended actions do not correct the situation, refer to the
BD Phoenix™ M50 User’s Manual or contact BD)
21
Error Codes W803 – W905