iMX113 BD Pyxis ™ ES Refrigerator Tower Operation Manual
360395/E 16
6 ProductSpecications
6.1 Operating Standards
These units are designed to operate under the following environmental conditions:
Non-residential, indoor use only
Altitude (maximum): 6.56k ft (2000 m)
Ambient temperature range: 59°F to 90°F (15°C to 32°C)
Relative humidity (maximum for ambient temperature): 80% for temperatures up to 89°F (31°C); 76% at 90°F (32°C)
Temperature control range: 36°F to 46°F (2°C to 8°C)
Overvoltage Category II
Pollution Degree 2
RF Emissions: Group 1 - Class A
EMC Environment: Basic
Sound level is less than 70 dB(A).
Table2.ElectricalSpecications
113
Input Voltage and Frequency 115V, 60Hz 230V 50/60Hz
Voltage Tolerance ±10%
Circuit Breakers 4.0A quantity 2
Current Draw* 2.3A 1.12A
Power Source Grounded outlet, meeting National Electric Code (NEC) in the U.S. and local electrical requirements in all locations.
Remote Alarm Capacity 1.0A at 33 V (RMS) or 30 V (DC)
* Amperage values are subject to change. Refer to the product specication label on your unit for current values.
NOTICE
The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
If an external power supply exceeding 33 V (RMS) or 30 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.