iMX113 BD Pyxis ™ ES Refrigerator Tower Operation Manual
360395/E 7
2 Installation
2.1 Location Requirements
Keep all ventilation openings in the enclosure or, in the structure for building-in, clear of obstruction.
Has a grounded outlet meeting the electrical requirements listed on the product specication label.
Is clear of direct sunlight, high temperature sources, and heating and air conditioning vents.
Minimum 8“ (203 mm) above, minimum 5” (127mm) right side, and minimum 3” (76 mm) behind for clearance and feature access.
Meets limits specied for ambient temperature (15˚C to 32˚C) and relative humidity.
2.2 Placement and Leveling
NOTICE
The evaporation tray located on the rear of the refrigerator may be hot. Do not use the tray as a handle. During normal
operation, evidence of evaporation such as visible steam and/or audible hiss may occur.
To prevent tipping, ensure the casters are unlocked, leveling feet (if installed) are lifted, and the doors are closed before
moving the refrigerator.
To avoid damaging refrigerant tubing or risking refrigerant leak, use caution when moving or operating the unit.
1. Remove refrigerator from shipping carton.
2. Remove and discard interior packing material.
3. Remove accessory package from inside refrigerator.
4. Ensure door is secured and casters are unlocked.
5. Roll refrigerator into place and lock casters.
6. Ensure refrigerator is level.
7. Ensure bins are locked in place within the modules.