BeaconMedaes HTM - Faults and Remedies: Controller

BeaconMedaes HTM
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4233 5000 13
79
Pressure sensors: error signal
(no measurement)
Electric connection
interrupted
Mechanical damage (e.g.
too high mounting torque)
Connection plugs not
connected
Crack due to incorrect
mounting
Internal defect
No supply voltage
Check connections
Check sensors
Pressure sensors: too high or
too low indication
Condense on connection
plugs
Faulty supply signal
Clogging or contamination
Sensor internally defect
Deviation due to ageing
Deviation due to incorrect
production
Check sensors
Replace defective parts
Pressure regulator: wrong
outlet pressure
Wrong setting
Check gauge of pressure regulator
Adjust pressure
Dew point sensor
malfunctioning
No direct feedback of dew point
Replace defective parts
Dew point sensor: no
measurement (error signal)
No direct feedback of dew point
Check connections
Check sensors
Dew point sensor: incorrect
measurement
Dew point alarm generated too
early or too late
Check sensors
Replace defective parts
Faults and remedies, controller
Condition Fault Remedy
Doesn't read in pressure or
dew point
Software fault
Reprogram the controller
Check wiring
Doesn’t issue alarm(s) when
it should
Software fault
Connection fault between
the controller and the
control room display
Reprogram the controller
Check wiring
Doesn't issue service warnings
when it should
Software fault Reprogram the controller
Keeps exhaust valves closed
(for purge of dryer)
Software fault
Communication fault
Reprogram the controller
Check wiring
Check solenoid