Cleaning the ejector
If error message "W 013" id displayed on the
screen, the so-called ejector must be cleaned.
1. Switch the device off and pull out the mains
plug.
2. Remove the cover (1) on the left side (loosen
the slotted screws).
3. Unscrew and remove the sound absorber (2)
and ejector.
4. Clean the ejector by blasting it.
5. The nozzle in the ejector must not be moved. It
must protrude 8 mm (see A).
Only use polish blasting agents that
do not remove material.
(recommended: BEGO Perlablast).