EasyManua.ls Logo

Benchmark TD-550 - Calibration Overview

Benchmark TD-550
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation TD-550/560 Oil in Water Analyzer
12 TD-550/560 Oil in Water Analyzer
Operation Manual
PN 106737
REV H
2.4 CALIBRATION OVERVIEW
Before the TD-550/560 can be used to analyze a sample, you must calibrate it with a blank and a
calibration standard containing your target hydrocarbon. If you are following a solvent-extraction
method, use the pure extraction solvent for the blank. Prepare the calibration standard by dissolving a
known amount of hydrocarbon in a known volume of extraction solvent. The standard solution will then
have a known concentration. You should also check the linearity and sensitivity as discussed in Section
6.1 Low Fluorescence During Calibration and Section 6.2 Linearity and Linearity Checking. Calibrations
are saved to a list on the Calibration screen. Each calibration includes a name, date, cuvette type, and
extraction ratio. On the Calibration screen, you can also associate a saved calibration with a location.
For fastHEX
TM
method calibration details, see Section 5 No-Solvent Analysis Method.
For No-Solvent method calibration details, see Section 5.2 Calibrating the Analyzer for 0-100 ppm
Range.
NOTE: Make sure that you use the same size cuvette for blank, calibration solution, and sample
analysis. For example, if you use the 8 mm cuvettes for blank and calibration solution, then you must
use 8 mm cuvettes for sample analysis also. Using different size cuvettes for calibration and sample
analysis will result in incorrect readings.