TD-550/560 Oil in Water Analyzer fastHEX™ Analysis Method
PN 106737
REV H
TD-550/560 Oil in Water Analyzer
Operation Manual
21
1
3
2
14. On the Home screen select a saved calibration in the Calibrations list (1).
NOTE: If you have not calibrated the analyzer, follow the calibration procedure in Section 5 No-
Solvent Analysis Method.
15. If applicable, select a saved location from the Locations list (2), or tap New to create a location.
16. Tap Read (3). A new reading is stored in the Reading History screen.
17. Check that the uorescence response is at least 200 uorescence units greater than the
blank. Calibrations with uorescence response differences less than 200 indicate insufcient
sensitivity. If this occurs, consult Section 6 Troubleshooting.
18. Dispose of the water, extraction solvent, cuvette and prescription bottle in a safe,
environmentally responsible manner, in accordance with your company’s procedures.
NOTE: For measurement of samples with expected concentrations near zero, see Section 6
Troubleshooting for suggestions on how to express the results.
For video tutorials on analyzing samples, see the How to Video Library on the Help & Settings
screen.