13.2 Cleaning, disinfection and sterilization
General cleaning and disinfection
The cleaning and disinfection of the operating light shall only be
made by a hygiene specialist or the person who is instructed by
the hygiene specialist.
Only chemicals which have been tested in terms of material
compatibility and released by BERCHTOLD shall be used. Special
care is necessary for the choice of the chemical product as well
as their concentration.
If the each current list, which can be view on the internet and is
also available on request (
back side of cover), does not include
a disinfectant or detergent, it shall not be used, since functional
components can be modified or damaged.
Prior to actual disinfection, perform a thorough cleaning to
remove any visible impurities, such as blood, etc.
Do not use any sharp, pointed or abrasive objects during
cleaning, as there is always a risk of damaging the surface.
Damaged surfaces become exposed to chemical substances
during subsequent cleaning or disinfection. This damages the
surface.
Therefore, only a soft brush and a mild cleaning agent (or a
cleaning disinfectant) are allowed for cleaning heavy and
stubborn impurities. When all visible impurities are gone,
disinfection is performed.
NOTE!
Any implied warranty claim terminates when these
guidelines are ignored.
NOTE!
The implied warranty is only valid for undamaged
surfaces!