30 WWW.BERTIN-INSTRUMENTS.COM
USER MANUAL
Use the Scroll up or Scroll down button to navigate and push the Enter button for
2 sec. to choose an operation mode.
Set operation mode
11.08.16 12:27:47 4min
60 min, diffusion, Radon
10 min, diffusion, Radon
10 min, flow, Radon
1 min, flow, Radon
10 min, flow, Radon/Thoron
Use ↑↓ to change and
press <┘key to accept !
Figure 20: Setting the operation mode
If the flow mode operation is chosen, the running pump is indicated by a pump
symbol in the upper right corner:
Figure 21: Acquisition window with pump symbol
7.2.4 Set pump flow
The integrated pump allows
radon and thoron measurements in flow mode. The
flow-regulated pump can be adjusted from flow rates of 0,05 l/min to 0,5, 1 and 2
l/min by using the Scroll down and Scroll up buttons. The adjustment is
acknowledged by pushing the Enter button for 2 sec.
- The contents of the display are updated along with each new measuring
cycle resp. every 10 min with the 60 min cycle.
-
A change of the measuring mode or measuring cycle automatically leads
to the creation of a new data series.
Note