LaserSpeed 8000/9000 I/O Module Instruction Handbook
Setup
Part No. 93342 / Drawing No. 0921-01516 Page 38 of 84 Revision E (Dec 2013)
Set Value
~B<B>,<S>,<D>,<P>,<M>,<E><CR>
.B<CR> Read Settings (without labels)
Example Response Format:
230400,1,8,NONE,0,10<CR>
Where <B>, <S>, <D>, <P>, <M> and <E> are as follows:
230400, 115200, 57600,
38400, 19200, 4800
Should always be set to
NONE
Number of elongation
readings to average
from 1 to 1000
Default 10
This command configures the baud rate and framing for the RS-422
communication link between the I/O Module and the LaserSpeed gauge. These
settings must be set to match the gauge’s communication settings for
communication between the two devices to function. The valid baud rates are
listed in the table above.
Recommended settings: