LaserSpeed 8000/9000 I/O Module Instruction Handbook
Setup
Part No. 93342 / Drawing No. 0921-01516 Page 43 of 84 Revision E (Dec 2013)
~F:UdpPort Ethernet UDP Port
~F:UdpPort<CR> Read Setting
Response Format:
ABEtherNet UdpPort <N><CR>
~F:UdpPort=<N><CR> Set Value
Factory Default: 4750 Minimum: 0 Maximum: 65535
Where <N> is the port on the host to which UDP messages will be sent. The
UDP message receiver should listen on this port for UDP messages.
This setting is only used when the I/O Module is equipped with an Ethernet
fieldbus card.
Example: To set the UDP Port 4750, send the command:
~F:UdpPort=4750<CR>
~F:TcpPort Ethernet TCP Port
~F:TcpPort<N> Read Setting
Response Format:
ABEtherNet TcpPort <N><CR>
~F:TcpPort=<N><CR> Set Value
Factory Default: 4751 Minimum: 0 Maximum: 65535
Where <N> is the local port for TCP configuration messages. The host computer
should open a TCP connection to this port in order to send configuration
messages.
This setting is only used when the I/O Module is equipped with an Ethernet
fieldbus card.
Example: To set the TCP Port 4751, send the command:
~F:TcpPort=4751<CR>