LaserSpeed 8000/9000 I/O Module Instruction Handbook
Operation
Part No. 93342 / Drawing No. 0921-01516 Page 73 of 84 Revision E (Dec 2013)
Profibus Status indicator
LED
Meaning
1
Not Used
2 (Green)
Indicates that the module is On-Line on the
fieldbus.
ON = Module is On-Line and data exchange is
possible.
OFF = Module is not On-Line
3 (Red)
Indicates that the module is Off-Line on the
fieldbus.
ON = Module is Off-Line and no data exchange is
possible.
OFF = Module is not Off-Line
4 (Red)
Fieldbus Diagnostics. Indicates certain faults on
the Fieldbus side.
Flashing Red 1 Hz - Error in configuration: IN
and/or OUT length set during initialisation of the
module is not equal to the length set during
configuration of the network.
Flashing Red 2 Hz - Error in User Parameter data:
The length/contents of the User Parameter data
set during initialisation of the module is not equal
to the length/contents set during configuration of
the network.
Flashing Red 4 Hz - Error in initialisation of the
Profibus communication ASIC.
Turned Off - No diagnostics present
There is also one additional Bicolour Watchdog LED on the AnyBus-S module.
The functionality of this LED is the same for all AnyBus-S modules, and is
specified in the AnyBus-S Design Guide.
1 2
3 4
Profibus Status Indicator