38 39
appears on the display. All this user’s values are
deleted.
The device switches o automatically.
7. CLEANING AND MAINTENANCE
• Clean the device and cu carefully using only a slightly
damp cloth.
• Do not use any cleaning solutions or solvents.
• Under no circumstances hold the device or cu under
water, as this can cause liquid to enter and damage the
device and cu.
• If you store the device and cu, do not place heavy ob-
jects on the device and cu. The cu line should not be
bent sharply.
• Remove the batteries if the device is not going to be
used for a long period of time.
8. ACCESSORIES AND
REPLACEMENT PARTS
Accessories and replacement parts are available on the
homepage www.beurer.de, under “Service”. Please state the
corresponding order number.
Designation Item number and/or
order number
Standard cu (22-36 cm) 163.388
XL cu (35-44 cm) 163.387
Mains part (EU) 071.95
9. TROUBLESHOOTING
Error
mes-
sage
Possible cause Solution
Unable to record
a pulse.
Please wait one minute
and repeat the meas-
urement.
Ensure that you do not
speak or move during
the measurement.
You moved or
spoke during the
measurement.
The cu was
not attached
correctly.
Please observe the
information in chapter
“Attaching the cu”.