Issue 05/2019 Art. No. 7001-0359
Operating Manual
BD
-S / BD-S-UL (E1)
Incubators with natural convection
ED
-S / ED-S-UL (E1)
Drying and heating ovens with natural
convection
FD
-S / FD-S-UL (E1)
Drying and heating ovens with forced
convection
with R-S microprocessor temperature controller
Model
Model version
Art. No.
BD-S 56 BDS056-230V 9090-0016, 9190-0016
BD-S-UL 56 BDS056UL-120V 9090-0017, 9190-0017
BD-S 115 BDS115-230V 9090-0022, 9190-0022
BD-S-UL 115 BDS115UL-120V 9090-0023, 9190-0023
ED-S 56 EDS056-230V 9090-0014, 9190-0014
ED-S-UL 56 EDS056UL-120V 9090-0015, 9190-0015
ED-S 115 EDS115-230V 9090-0020, 9190-0020
ED-S-UL 115 EDS115UL-120V 9090-0021, 9190-0021
FD-S 56 FDS056-230V 9090-0018, 9190-0018
FD-S-UL 56 FDS056UL-120V 9090-0019, 9190-0019
FD-S 115 FDS115-230V 9090-0024, 9190-0024
FD-S-UL 115 FDS115UL-120V 9090-0025, 9190-0025
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • E
    Erica RodriguezAug 14, 2025
    What to do if the Binder BD-S 56 Oven chamber doesn’t heat up and the 'Safety controller active' LED is on?
    • A
      Alejandra RogersAug 14, 2025
      If the Binder Oven chamber doesn't heat up and the 'Safety controller active' LED is lit, it could be due to a few reasons: * For BD-S / BD-S-UL models: The safety controller class 3.1 might have been activated. Check the temperature set-point and safety controller settings, and select an appropriate limit value. * For ED-S / ED-S-UL, FD-S / FD-S-UL models: The safety controller class 2 may have turned off the heating. Allow the chamber to cool down, then press the confirmation button. Also, verify the temperature set-point and safety controller settings, adjusting the limit value if needed. * The safety controller itself might be defective, in which case you should contact BINDER service.
  • B
    bbrooksAug 17, 2025
    Why Binder Oven set-point temperature is not reached after specified time?
    • H
      hjohnsonAug 17, 2025
      If your Binder Oven's set-point temperature isn't reached after the specified time, consider the following possible causes: * The chamber door may not be properly closed; ensure it is completely shut. * The door gasket might be defective and need replacement. * The controller may require calibration and adjustment. * The power supply might be providing the wrong voltage; verify that it matches the oven's requirements.
  • E
    Ernest ThompsonAug 21, 2025
    What to do if Binder BD-S 56 Oven chamber doesn’t heat up but the 'Heating active' LED is lit?
    • L
      Lee WoodsAug 21, 2025
      If your Binder Oven chamber isn't heating up, but the 'Heating active' LED is on, it could be due to: * A defective heating element, in which case you should contact BINDER service. * A defective semiconductor relay.
  • A
    Amanda PerryAug 25, 2025
    Why Binder BD-S 56 chamber is heating permanently, and the set-point is not held?
    • A
      Alexander MillsAug 25, 2025
      If the Binder Oven chamber heats permanently and the set-point temperature isn't maintained, it could be due to: * A defective controller, in which case you should contact BINDER service. * A defective Pt 100 sensor. * A defective semiconductor relay. * The controller may need calibration and adjustment.
  • D
    Darrell PrestonAug 27, 2025
    What to do if Binder Oven chamber doesn’t heat up and the 'Heating active' LED is not lit?
    • D
      Dana MeyerAug 27, 2025
      If the Binder Oven chamber isn't heating up, the 'Heating active' LED is off, but the controller display is working, it could be due to: * The timer running off, so reprogram the timer or turn it off. * A defective semiconductor relay, in which case you should contact BINDER service. * A defective controller.
  • D
    Daisy TuckerAug 30, 2025
    How to troubleshoot Binder BD-S 56 chamber without function and dark display?
    • M
      Michael LiuAug 30, 2025
      If the Binder Oven chamber has no function and the controller display is dark, it could be caused by: * No power supply; check the connection to the power supply. * The overtemperature protective device class 1 has turned off the chamber, in which case you should contact BINDER service. * A defective controller, in which case you should contact BINDER service.
  • J
    Justin McgrathSep 2, 2025
    What to do if Binder BD-S 56 controller shows “OFF” on lower display?
    • A
      Abigail ZunigaSep 3, 2025
      If the controller shows “OFF” on the lower display and the actual temperature value on the upper display, the chamber is likely in standby mode. To resolve this, simply press the confirmation button.
  • M
    Miss Ashley SpencerSep 6, 2025
    What to do if FD-S / FD-S-UL Binder BD-S 56 Oven fan doesn’t turn or turns too slowly?
    • T
      Theodore WilliamsSep 6, 2025
      If the FD-S / FD-S-UL Binder Oven fan doesn’t turn or turns too slowly, the fan may be defective, in which case you should contact BINDER service.
  • S
    Sharon MooreSep 9, 2025
    Why does Binder BD-S 56 Oven have deviations from the indicated heating-up times?
    • C
      Chelsea LewisSep 9, 2025
      If you are experiencing deviations from the indicated heating-up times with your Binder Oven, it may be due to the chamber being fully loaded. Try loading the chamber with fewer items or allow for longer heating-up times.

Summary

1. Safety

Legal Considerations

Explains general legal and informational aspects of the manual.

Structure of Safety Instructions

Details the organization and symbols of safety information.

2. Chamber Description

Chamber Overview

Provides a visual overview of the chamber components with labels.

Triangular Instrument Panel

Describes the triangular instrument panel and controller functions.

3. Completeness of Delivery, Transportation, Storage, and Installation

Unpacking and Delivery Checks

Details the process of unpacking and checking the chamber upon delivery.

Safe Lifting and Transportation

Covers safe lifting and transportation guidelines for the chamber.

Storage and Ambient Conditions

Provides guidelines for storing the chamber and acceptable ambient conditions.

4. Installation

Installing the Racks

Instructions for inserting rack supports and racks into the chamber.

Ventilation System Connection

Guides on connecting optional exhaust or ventilation systems.

Electrical Connection

Details the electrical connection requirements and plug types.

5. R-S Controller Overview

Controller Display and Buttons

Describes the R-S controller's display elements and functional buttons.

Menu Structure and Navigation

Explains the controller's menu structure, quick access, and password-protected menus.

6. Start Up

Initial Startup Procedure

Steps to start up the chamber, including plugging in and checking the controller.

Adjusting Air Change

Instructions on how to adjust the air change using the exhaust air flap.

7. Temperature Set-Point Entry

Entering Temperature Set-Points

Explains how to enter and confirm the desired temperature set-point.

8. Selecting the Temperature Unit

Setting the Temperature Unit

Guides on changing the temperature display between Celsius and Fahrenheit.

9. Overtemperature Protection

Class 1 Protective Device

Describes the function and limitations of the class 1 safety device.

Safety Controller (Class 2;3.1)

Explains the safety controller operation, settings, and alarm procedures.

10. Timer Function Delayed Off

Setting Timer Run-Time

Instructions for setting the delayed off timer function and its run-time.

11. Ramp Function

Setting the Ramp Gradient

Guides on setting the ramp gradient and target temperature for controlled heating.

Displaying and Turning Off Ramp

Explains how to view the ramp course and how to turn off the ramp function.

12. Options

Data Logger Kits

Information about optional data logger kits for temperature monitoring.

13. Maintenance, Cleaning, and Service

Maintenance Intervals and Service

Outlines recommended maintenance intervals and service procedures.

Cleaning Procedures

Detailed steps for cleaning interior and exterior surfaces.

Decontamination Procedures

Steps for decontaminating the chamber after exposure to hazardous substances.

Returning Chamber for Repair

Instructions and requirements for sending the chamber back to BINDER for service.

14. Disposal

Transport Packing Disposal and Decommissioning

Guidance on disposing of packing materials and decommissioning the chamber.

Chamber Disposal in Germany and EU

Procedures for disposing of the chamber in Germany and other EU member states.

Chamber Disposal Outside the EU

Procedures for disposing of the chamber in non-member states of the EU.

15. Troubleshooting

Troubleshooting Common Issues

Lists common faults, their possible causes, and required corrective measures.

16. Technical Description

Factory Calibration and Usable Volume Definition

Details factory calibration processes and the definition of usable volume.

Over Current Protection and Technical Data

Explains over current protection and provides detailed technical specifications.

Equipment, Options, Spares, and Drawings

Lists standard equipment, optional accessories, spare parts, and dimensioned drawings.

17. Certificates and Declarations of Conformity

EU Declarations of Conformity

Presents EU conformity declarations for different chamber models.

VDE and UL Compliance Certificates

Provides GS mark certificate from VDE and UL compliance certificate.

18. Product Registration

Online Product Registration

Guides users on how to register their BINDER product online.

19. Contamination Clearance Certificate

Contamination Certificate (Outside USA;Canada)

Form for declaring safety and health information for returns outside USA/Canada.

Product Return & Decontamination (USA;Canada)

Request form for returns and decontamination declaration for USA/Canada.

Binder BD-S 56 Specifications

General IconGeneral
ModelBD-S 56
Power Supply230 V, 50/60 Hz
Temperature Range5 °C above ambient temperature to 300 °C
Interior Volume56 L
Number of Shelves (std./max.)2 /
Timer0 to 99 h 59 min
DoorInner glass door