BD-S / BD-S-UL, ED-S / ED-S-UL, FD-S / FD-S-UL (E1) 05/2019 page 19/59
8. Selecting the temperature unit
You can set the temperature unit to degrees Celsius °C or degrees Fahrenheit °F.
If the unit is changed, all temperature values are converted accordingly.
Also when specifying the ramp function (see chap. 11) this setting is accordingly taken as the basis.
C = degrees Celsius
F= degrees Fahrenheit
0 °C = 31°F
100 °C = 212°F
Conversion:
[Value in °F] = [Value in °C] 1,8 + 32
8.1 Setting the temperature unit
Simultaneously press the confirmation button and the arrow-up button to ac-
cess the „Setup“ menu. The password entry display “S.Locflashes.
Enter the password 10with the arrow-up button and confirm with the confirma-
tion button.
Press the arrow-up button to access the “Unit“ menu (temperature unit). The
„Unit“ menu is shown with the current temperature unit.
Press the confirmation button to activate the entry. The display “Unit” flashes.
Select the desired unit with the arrow buttons and confirm with the confirmation
button.
Keep pressed down the confirmation button and press the arrow-up button to
return to Normal display.
Example:
+
10x
or
9. Overtemperature protection
9.1 Overtemperature protective device (class 1)
The chambers are equipped with an internal temperature safety device, class 1 acc. to DIN 12880:2007.
It serves to protect the unit and prevents dangerous conditions caused by major defects.
If the cut-off temperature is reached, the over temperature protective device permanently turns off the
unit. The user cannot restart the device again. The protective cut-off device is located internally. Only a
service specialist can replace it. Therefore, please contact an authorized service provider or BINDER
service.
Cut-off temperature values:
BD-S / BD-S-UL: 90 °C / 194 °F
ED-S / ED-S-UL, FD-S / FD-S-UL: 318 °C / 604.4 °F