Issue 03/2019 Art. No. 7001-0026
Operating Manual
Translation of the original operating manual
BD / BD-UL (E2) Incubators with natural convection
ED / ED-UL (E2) Drying and heating ovens with natural convection
FD / FD-UL (E2) Drying and heating ovens with forced convection
with microprocessor temperature controller
Modell Model version Art. No. Modell Model version Art. No.
BD 23
BD023-230V 9010-0187
ED 240
ED240-230V 9010-0101
BD023UL-120V 9010-0189 ED240UL-208V 9010-0167
BD 53
BD053-230V 9010-0081
ED400
ED400-230V 9010-0075
BD053UL-120V 9010-0179 ED400UL-208V 9010-0168
BD 115
BD115-230V 9010-0088
ED720
ED720-230V 9010-0076
BD115UL-120V 9010-0181 ED720UL-208V 9010-0169
BD 240
BD240-230V 9010-0095
FD 23
FD023-230V 9010-0194
BD240UL-120V 9010-0183 FD023UL-120V 9010-0196
BD400
BD400-230V 9010-0073
FD 53
FD053-230V 9010-0082
BD400UL-120V 9010-0176 FD053UL-120V 9010-0128
BD720
BD720-230V 9010-0074
FD 115
FD115-230V 9010-0102
BD720UL-120V 9010-0177 FD115UL-120V 9010-0129
ED 23
ED023-230V 9010-0190/0191
FD 240
FD240-230V 9010-0104
ED023UL-120V 9010-0192/0193 FD240UL-208V 9010-0130
ED 53
ED053-230V 9010-0078/0079
ED053UL-120V 9010-0131/0132
ED 115
ED115-230V 9010-0096/0097
ED115UL-120V 9010-0164/0165
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • J
    Jade SimmonsAug 22, 2025
    Why is my Binder Climate chamber not working?
    • W
      William ThompsonAug 22, 2025
      Your Binder Climate chamber might not be working due to several reasons. First, check the connection to the power supply. Second, the chamber fuse might have responded, so check it and replace it if needed. If it responds again, contact BINDER service. Finally, the controller could be defective, in which case you should also contact BINDER service.
  • G
    Gabriel HortonSep 4, 2025
    Why does my Binder ED 23 Climate chamber controller return to normal display?
    • C
      Christine SandersSep 4, 2025
      If the controller on your Binder Climate chamber returns to the Normal Display from any level, it's likely because no button was pressed for approximately 30 seconds. Repeat your entries, entering the values more rapidly.
  • S
    sarah35Sep 3, 2025
    What to do if Binder ED 23 fan is not working?
    • E
      echarlesSep 3, 2025
      If the fan in your Binder Climate chamber isn’t turning or is turning too slowly, the fan might be defective, in that case contact BINDER service.
  • A
    Adrian BishopSep 1, 2025
    Why is my Binder Climate chamber in stand-by mode?
    • E
      Eileen WarrenSep 1, 2025
      If your Binder Climate chamber is without function and only the green 'stand-by' LED is lit, the chamber is likely in stand-by mode. Press the ON/OFF button until the display lights up.
  • J
    Jordan MorenoAug 31, 2025
    Why does my Binder ED 23 take longer to heat up?
    • D
      David HarperAug 31, 2025
      If you're experiencing deviations from the indicated heating-up times with your Binder Climate chamber, it might be because the chamber is fully loaded. Try loading the chamber with fewer items or allow for longer heating-up times.
  • N
    nwilsonAug 29, 2025
    How to fix high temperature and red alarm on Binder ED 23 Climate chamber?
    • A
      Andrea WebsterAug 29, 2025
      If the temperature inside your Binder Climate chamber is too high and the red alarm light is on, the safety device class 3.1 may have been triggered. Check the settings of the temperature set-point and of the safety device class 3.1.
  • R
    Richard OlsonAug 27, 2025
    Why is my Binder ED 23 Climate chamber heating permanently?
    • I
      ibernardAug 28, 2025
      If your Binder Climate chamber is heating permanently and not maintaining the set-point, the controller may be defective, in that case contact BINDER service. Alternatively, the controller might need calibration and adjustment.
  • R
    robertprattAug 27, 2025
    What to do if Binder ED 23 Climate chamber isn't working and the red alarm light is on?
    • D
      debracampbellAug 27, 2025
      If your Binder Climate chamber isn't working and the red alarm pilot lamp is lit, it could be that the safety device class 2 has turned off the chamber. Let the chamber cool down and then press the RESET button. Also, the safety device class 2 might be defective, in that case contact BINDER service.
  • J
    Jon BlackburnAug 25, 2025
    Why Binder ED 23 Climate chamber doesn’t heat up but controller display working?
    • M
      Mark CollinsAug 25, 2025
      If your Binder Climate chamber isn’t heating up, and the red heating control light is off but the controller display is working, it could be due to the timer being run off. Program the timer or switch to continuous operation. Another cause might be a defective semiconductor relay, in which case you should contact BINDER service.
  • K
    Kayla JacksonAug 23, 2025
    What to do if Binder ED 23 Climate chamber set-point temperature is not reached?
    • J
      Jason NguyenAug 23, 2025
      If the set-point temperature isn't reached in your Binder Climate chamber, ensure the chamber door is completely closed. Also, inspect the door gasket and replace it if it’s defective. Calibrate and adjust the controller. Finally, check the power supply to ensure the voltage is correct (115V or 230V).

Summary

Safety and Compliance

Safety Instructions and Warnings

Explains safety symbols, labels, and general precautions for safe operation.

Intended Use and Accident Prevention

Defines permissible applications and manufacturer's measures to prevent accidents.

Chamber Installation and Setup

Chamber Description and Overview

General description of chamber components, materials, and key features.

Unpacking, Transportation, and Storage

Guides on initial handling, movement, and intermediate storage of the chamber.

Installation Location and Electrical Connection

Specifies site requirements and electrical hook-up procedures for proper installation.

Controller Operation and Settings

Temperature Control and Ramps

Instructions for setting temperature set-points and programming temperature ramps.

Time Functions and User Settings

Explains timer operation, continuous modes, and user-configurable settings.

Safety Devices and Options

Temperature Safety Device Functions

Information on Class 2 and Class 3.1 safety devices for temperature protection.

Optional Features and Accessories

Overview of optional features like data loggers, sensors, and alarms.

Maintenance, Cleaning, and Disposal

Maintenance, Cleaning, and Decontamination

Guidelines for regular maintenance, cleaning, and decontamination procedures.

Chamber Disposal Guidelines

Procedures for environmentally friendly disposal of the chamber in various regions.

Troubleshooting and Technical Data

Troubleshooting Common Issues

Solutions for common problems encountered during chamber operation.

Technical Specifications (BD;ED;FD)

Detailed technical data and specifications for different chamber models.

Certificates and Product Registration

Conformity Declarations and Certificates

Official conformity declarations and safety mark certificates for the equipment.

Online Product Registration

Instructions for registering the BINDER product online for service benefits.

Contamination Clearance for Returns

Return Procedures and Decontamination Forms

Forms and procedures for returning equipment, including decontamination requirements.

Binder ED 23 Specifications

General IconGeneral
BrandBinder
ModelED 23
CategoryClimate chamber
LanguageEnglish

Related product manuals