Issue 03/2019 Art. No. 7001-0026
Operating Manual
Translation of the original operating manual
BD / BD-UL (E2) Incubators with natural convection
ED / ED-UL (E2) Drying and heating ovens with natural convection
FD / FD-UL (E2) Drying and heating ovens with forced convection
with microprocessor temperature controller
Modell Model version Art. No. Modell Model version Art. No.
BD 23
BD023-230V 9010-0187
ED 240
ED240-230V 9010-0101
BD023UL-120V 9010-0189 ED240UL-208V 9010-0167
BD 53
BD053-230V 9010-0081
ED400
ED400-230V 9010-0075
BD053UL-120V 9010-0179 ED400UL-208V 9010-0168
BD 115
BD115-230V 9010-0088
ED720
ED720-230V 9010-0076
BD115UL-120V 9010-0181 ED720UL-208V 9010-0169
BD 240
BD240-230V 9010-0095
FD 23
FD023-230V 9010-0194
BD240UL-120V 9010-0183 FD023UL-120V 9010-0196
BD400
BD400-230V 9010-0073
FD 53
FD053-230V 9010-0082
BD400UL-120V 9010-0176 FD053UL-120V 9010-0128
BD720
BD720-230V 9010-0074
FD 115
FD115-230V 9010-0102
BD720UL-120V 9010-0177 FD115UL-120V 9010-0129
ED 23
ED023-230V 9010-0190/0191
FD 240
FD240-230V 9010-0104
ED023UL-120V 9010-0192/0193 FD240UL-208V 9010-0130
ED 53
ED053-230V 9010-0078/0079
ED053UL-120V 9010-0131/0132
ED 115
ED115-230V 9010-0096/0097
ED115UL-120V 9010-0164/0165
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • M
    Mary DavidsonSep 8, 2025
    What to do if the Binder Accessories chamber doesn't heat up, but the red heating control light is not lit and the controller display is working?
    • C
      Christopher WilkinsSep 8, 2025
      If your Binder Accessories chamber isn't heating up, but the red heating control light in the display is off and the controller display is working, consider the following: * **Timer:** Check if the timer has run out. Program the timer or switch to continuous operation mode as described in chapter 6.3. * **Semiconductor Relay:** It is possible that the semiconductor relay is defective, in which case you should contact BINDER service.
  • I
    icookAug 23, 2025
    What to do if Binder BD 23 Climate chamber set-point temperature is not reached?
    • T
      tyuAug 23, 2025
      If the set-point temperature isn't reached in your Binder Climate chamber, ensure the chamber door is completely closed. Also, inspect the door gasket and replace it if it’s defective. Calibrate and adjust the controller. Finally, check the power supply to ensure the voltage is correct (115V or 230V).
  • J
    joelcortezAug 25, 2025
    Why Binder BD 23 Climate chamber doesn’t heat up but controller display working?
    • A
      Angel HarrisonAug 26, 2025
      If your Binder Climate chamber isn’t heating up, and the red heating control light is off but the controller display is working, it could be due to the timer being run off. Program the timer or switch to continuous operation. Another cause might be a defective semiconductor relay, in which case you should contact BINDER service.
  • R
    Renee WrightAug 26, 2025
    What to do if Binder BD 23 isn't working and the red alarm light is on?
    • M
      Maureen SmithAug 26, 2025
      If your Binder Climate chamber isn't working and the red alarm pilot lamp is lit, it could be that the safety device class 2 has turned off the chamber. Let the chamber cool down and then press the RESET button. Also, the safety device class 2 might be defective, in that case contact BINDER service.
  • J
    jessicataylorAug 27, 2025
    Why is my Binder BD 23 heating permanently?
    • M
      miranda04Aug 27, 2025
      If your Binder Climate chamber is heating permanently and not maintaining the set-point, the controller may be defective, in that case contact BINDER service. Alternatively, the controller might need calibration and adjustment.
  • J
    Jessica DavisSep 22, 2025
    Why does the Binder Accessories controller return to Normal Display?
    • R
      rphillipsSep 22, 2025
      The controller returns to the Normal Display if no button is pressed for approximately 30 seconds. To prevent this, repeat your entries and enter the values quickly.
  • L
    Linda RodgersSep 16, 2025
    What to do if the Binder Accessories fan doesn’t turn or turns too slowly?
    • M
      Mark BryanSep 16, 2025
      If the fan in your Binder Accessories chamber doesn’t turn or turns too slowly, the fan might be defective. Contact BINDER service.
  • A
    Amy WilliamsSep 18, 2025
    Why is my Binder Accessories chamber heating permanently and not holding the set-point?
    • A
      alfred78Sep 18, 2025
      If your Binder Accessories chamber is heating permanently and not maintaining the set-point, it could be due to several reasons: * **Controller:** The controller may be defective. In this case, contact BINDER service. * **Controller:** The controller may need calibration. Calibrate and adjust the controller.
  • E
    Eric BrownSep 7, 2025
    What to do if my Binder Accessories chamber doesn't work?
    • A
      Adam JacksonSep 7, 2025
      If your Binder Accessories chamber is not functioning, here are a few things to check: * **Power Supply:** Ensure the chamber is properly connected to a power source. * **Voltage:** Verify that the power supply provides the correct voltage. Refer to chapter 4.1 for specifications. * **Chamber Fuse:** Inspect the chamber fuse and replace it if necessary. If the fuse fails again, it is recommended to contact BINDER service. * **Controller:** If the above steps don't resolve the issue, the controller might be defective, and you should contact BINDER service.
  • J
    justin85Sep 4, 2025
    Why does my Binder BD 23 Climate chamber controller return to normal display?
    • K
      Kimberly ChavezSep 4, 2025
      If the controller on your Binder Climate chamber returns to the Normal Display from any level, it's likely because no button was pressed for approximately 30 seconds. Repeat your entries, entering the values more rapidly.

Summary

Safety

General Safety Instructions

Intended Use

Accident Prevention Measures

Chamber Description

Delivery, Transportation, Storage & Installation

Safe Lifting and Transportation

Procedures for safely moving and transporting the chamber.

Installation Location and Ambient Conditions

Requirements for the installation site and environmental parameters.

Equipment Installation

Electrical Connection

Details on connecting the chamber to the power supply.

Start Up

Turning on the Chamber

Procedures for powering on the chamber via the main switch and controller.

Controller Operation

Temperature Set-Point Entry (No Ramp)

How to set target temperatures without using ramp functions.

Temperature Set-Point Entry (With Ramp)

How to set target temperatures using programmed temperature ramps.

Timer Operation: Setting Time

How to set and manage the timer for specific operation durations.

User Level Settings

Entering a Temperature Ramp

Programming temperature ramps for controlled heating profiles.

Temperature Safety Devices

Safety Device Class 2 (ED, FD)

Details on the Class 2 safety device for ED and FD models.

Safety Device Class 3.1 (BD)

Details on the Class 3.1 safety device for BD models.

Maintenance, Cleaning, and Service

Troubleshooting

Technical Description

BD Technical Data

Comprehensive technical specifications for BD series chambers.

Binder BD 23 Specifications

General IconGeneral
Humidity Range10% to 80% RH
Power Supply230 V, 50/60 Hz
Number of Shelves (standard/max)2/4
Max. Total Load30 kg
Volume23 liters
Temperature Fluctuation at 25°C±0.3 K