Issue 07/2017 Art. No. 7001-0130
Operating Manual
FP – Drying and heating ovens
with forced convection and program functions
with microprocessor program controller RD3
Model Model version Art. No.
FP 53 (E1.1) FP053-230V 9010-0153, 9110-0153
FP 53-UL (E1.1) FP053UL-120V 9010-0175, 9110-0175
FP 115 (E1.1) FP115-230V 9010-0255, 9110-0255
FP 115-UL (E1.1) FP115UL-120V 9010-0262, 9110-0262
FP 240 (E1.1) FP240-230V 9010-0263, 9110-0263
FP 240-UL (E1.1) FP240UL-120V 9010-0264, 9110-0264
FP 400 (E1.1) FP400-230V 9010-0265, 9110-0265
FP 400-UL (E1.1) FP400UL-120V 9010-0266, 9110-0266
FP 720 (E1.1) FP720-230V 9010-0267, 9110-0267
FP 720-UL (E1.1) FP720UL-120V 9010-0268, 9110-0268
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • K
    Kevin StewartAug 17, 2025
    What to do if the set-point temperature is not reached in my Binder Laboratory Equipment?
    • H
      howesamuelAug 17, 2025
      If the set-point temperature isn't reached after the specified time, there might be a few reasons. First, make sure the chamber door is completely closed. If it is, inspect the door gasket for any defects and replace it if necessary. Also, the controller might need adjustment; try calibrating it.
  • M
    Mark ThompsonAug 20, 2025
    What causes my Binder FP 240 Laboratory Equipment chamber to heat permanently and not hold the set-point?
    • M
      Mark HayesAug 20, 2025
      If your chamber is heating permanently and not holding the set-point, calibrate and adjust the controller. For other potential causes, contact Binder service.
  • M
    Margaret MooreAug 23, 2025
    Why are there deviations from the indicated heating-up times in my Binder Laboratory Equipment?
    • L
      lisamartinezAug 23, 2025
      Deviations from the indicated heating-up times can occur if the oven is fully loaded. Try charging the oven with less material or consider allowing longer heating-up times.
  • X
    xdavisAug 25, 2025
    Why is my Binder FP 240 chamber permanently turned off?
    • E
      ebrownAug 26, 2025
      If the chamber is permanently turned off, first check the connection to the power supply. Ensure the correct voltage (115V or 230V) is being supplied. Also, check the chamber fuse and replace it if needed. If the fuse responds again, contact Binder service.
  • D
    Doris McdanielAug 27, 2025
    What to do if the red indicator lamp is lit and my Binder Laboratory Equipment chamber is permanently turned off?
    • R
      Ronald SmithAug 28, 2025
      If the red indicator lamp of safety device class 2 is lit and the chamber is permanently turned off, let the oven cool down and then press the RESET button. Check the temperature set-point and the setting of the safety device, and select a suitable limit value if appropriate.
  • K
    Kenneth Rodriguez MDAug 30, 2025
    What to do if my Binder Laboratory Equipment chamber doesn’t heat up, the LED is not lit, but the controller display is working?
    • M
      Mrs. Alyssa FrancoAug 30, 2025
      If the chamber doesn’t heat up, the LED is not lit, and the controller display is working, contact Binder service.
  • Y
    yshermanSep 1, 2025
    What to do if my Binder FP 240 chamber doesn’t heat up, but the LED is lit?
    • M
      mcbrideaaronSep 2, 2025
      If the chamber doesn’t heat up but the LED is lit, contact Binder service.
  • J
    Julie ScottSep 4, 2025
    Why is the program duration longer than programmed on my Binder FP 240 Laboratory Equipment?
    • S
      Sarah MaciasSep 4, 2025
      If the program duration is longer than programmed, it may be due to inappropriate tolerances. For rapid transition phases, do NOT program tolerance limits to allow maximum heating speed.
  • J
    Jacob DavisSep 6, 2025
    How to fix ramp temperature transitions that are only realized as steps in my Binder FP 240?
    • S
      Steve GuerreroSep 7, 2025
      If ramp temperature transitions are only realized as steps, set the set-point programming type to “Ramp” in the User level.
  • D
    Daniel DixonSep 9, 2025
    Why does my Binder FP 240 program stop one section too early?
    • D
      Derrick StevensonSep 9, 2025
      If the program stops one section too early, the program line might be incomplete. Define the end value of the desired cycle by adding an additional section with a section time of at least one minute when programming.

Summary

Safety

Structure of Safety Instructions

Harmonized denominations, symbols, signal words, and pictograms for dangerous situations.

General Safety Instructions

Guidance on safe installation, operation, and usage of BINDER chambers and accessories.

Scope of Delivery, Transportation, Storage, and Installation

Installation of Equipment

Electrical Connection

Details on power connection, plugs, voltage, and electrical safety requirements.

Start Up Procedures

RD3 Program Controller Settings

Initial settings and display indications for the RD3 controller upon chamber startup.

Fixed Value Entry Mode

Week Program Editor

Program Editor

Set-point Ramp vs. Step Programming

Selecting between 'Ramp' and 'Step' for temperature transitions and its implications.

Set-point Entry for Program Operation

Procedure for entering set-points for program sections via the program editor.

Program Start Level

Program Selection and Course Settings

Selecting a program and defining parameters like delay time and number of cycles.

User Level Settings

Controller Configuration

Settings for menu language, program types, section duration, and set-point programming.

Example Programming: Week Program Editor

Example Programming: Program Editor

Detailed Programming Procedure

Step-by-step instructions for entering program data into the program editor.

Behavior at Failures

Alarm Messages

Explanation of displayed alarm messages and buzzer activation/reset.

Temperature Safety Devices

Safety Device Class 2 (DIN 12880)

Function, setting, and checking of the Class 2 temperature safety device.

Safety Device Class 3.1 (DIN 12880) (Option)

Function, setting, and checking of the optional Class 3.1 temperature safety device.

Maintenance, Cleaning, and Service

Cleaning Procedures

Detailed instructions for cleaning the chamber interior, exterior, and components with recommended agents.

Decontamination Procedures

Steps for decontaminating the chamber after exposure to hazardous substances.

Disposal Procedures

Chamber Disposal in Germany

Procedure for disposing of the chamber in Germany according to WEEE regulations.

Chamber Disposal in EU Member States (Excluding Germany)

Procedure for disposing of the chamber in other EU states according to WEEE regulations.

Troubleshooting Guide

Heating Issues

Common faults related to heating performance and their possible causes and solutions.

Controller Issues

Troubleshooting common faults related to the program controller's operation and settings.

Technical Description

Technical Data Tables

Detailed specifications including dimensions, capacity, temperature, and electrical data for all chamber sizes.

Binder FP 240 Specifications

General IconGeneral
Interior Volume240 L
Number of Shelves2
Max. Shelf Load30 kg
Voltage230 V
Frequency50/60 Hz
Temperature Range5 °C above ambient temperature to 300 °C
Power Consumption3.0 kW