Issue 07/2017 Art. No. 7001-0130
Operating Manual
FP – Drying and heating ovens
with forced convection and program functions
with microprocessor program controller RD3
Model Model version Art. No.
FP 53 (E1.1) FP053-230V 9010-0153, 9110-0153
FP 53-UL (E1.1) FP053UL-120V 9010-0175, 9110-0175
FP 115 (E1.1) FP115-230V 9010-0255, 9110-0255
FP 115-UL (E1.1) FP115UL-120V 9010-0262, 9110-0262
FP 240 (E1.1) FP240-230V 9010-0263, 9110-0263
FP 240-UL (E1.1) FP240UL-120V 9010-0264, 9110-0264
FP 400 (E1.1) FP400-230V 9010-0265, 9110-0265
FP 400-UL (E1.1) FP400UL-120V 9010-0266, 9110-0266
FP 720 (E1.1) FP720-230V 9010-0267, 9110-0267
FP 720-UL (E1.1) FP720UL-120V 9010-0268, 9110-0268
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: service@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Questions and Answers

  • P
    Patricia JonesAug 18, 2025
    What to do if the set-point temperature is not reached in my Binder FP 115 Laboratory Equipment?
    • S
      sbrennanAug 18, 2025
      If the set-point temperature isn't reached after the specified time, there might be a few reasons. First, make sure the chamber door is completely closed. If it is, inspect the door gasket for any defects and replace it if necessary. Also, the controller might need adjustment; try calibrating it.
  • T
    Tammy BuckAug 20, 2025
    What causes my Binder FP 115 Laboratory Equipment chamber to heat permanently and not hold the set-point?
    • P
      pattersontamaraAug 20, 2025
      If your chamber is heating permanently and not holding the set-point, calibrate and adjust the controller. For other potential causes, contact Binder service.
  • R
    ricardo76Aug 22, 2025
    Why are there deviations from the indicated heating-up times in my Binder FP 115 Laboratory Equipment?
    • M
      Mr. David KhanAug 23, 2025
      Deviations from the indicated heating-up times can occur if the oven is fully loaded. Try charging the oven with less material or consider allowing longer heating-up times.
  • H
    Hannah Evans MDAug 26, 2025
    Why is my Binder FP 115 Laboratory Equipment chamber permanently turned off?
    • S
      Steven WhiteheadAug 26, 2025
      If the chamber is permanently turned off, first check the connection to the power supply. Ensure the correct voltage (115V or 230V) is being supplied. Also, check the chamber fuse and replace it if needed. If the fuse responds again, contact Binder service.
  • R
    Rebekah HerreraAug 27, 2025
    What to do if the red indicator lamp is lit and my Binder FP 115 Laboratory Equipment chamber is permanently turned off?
    • H
      Heidi LeeAug 27, 2025
      If the red indicator lamp of safety device class 2 is lit and the chamber is permanently turned off, let the oven cool down and then press the RESET button. Check the temperature set-point and the setting of the safety device, and select a suitable limit value if appropriate.
  • S
    Samantha Simmons DVMAug 30, 2025
    What to do if my Binder FP 115 chamber doesn’t heat up, the LED is not lit, but the controller display is working?
    • V
      Veronica HarmonAug 30, 2025
      If the chamber doesn’t heat up, the LED is not lit, and the controller display is working, contact Binder service.
  • M
    mandersonSep 2, 2025
    What to do if my Binder Laboratory Equipment chamber doesn’t heat up, but the LED is lit?
    • C
      csanchezSep 2, 2025
      If the chamber doesn’t heat up but the LED is lit, contact Binder service.
  • J
    Joshua WoodsSep 4, 2025
    Why is the program duration longer than programmed on my Binder FP 115?
    • B
      Brent FernandezSep 4, 2025
      If the program duration is longer than programmed, it may be due to inappropriate tolerances. For rapid transition phases, do NOT program tolerance limits to allow maximum heating speed.
  • J
    jack71Sep 7, 2025
    How to fix ramp temperature transitions that are only realized as steps in my Binder Laboratory Equipment?
    • B
      Benjamin FrancoSep 7, 2025
      If ramp temperature transitions are only realized as steps, set the set-point programming type to “Ramp” in the User level.
  • M
    Melanie Alvarez DDSSep 9, 2025
    Why does my Binder Laboratory Equipment program stop one section too early?
    • J
      joseph46Sep 9, 2025
      If the program stops one section too early, the program line might be incomplete. Define the end value of the desired cycle by adding an additional section with a section time of at least one minute when programming.

Summary

FP Series Oven Overview

Safety Information

Installation and Setup

Unpacking, Transportation, and Storage

Instructions for unpacking, safe lifting, transport, and storage of the chamber.

Location and Ambient Conditions

Requirements for chamber placement, ventilation, and ambient operating conditions.

Electrical Connection

Start Up Procedures

Air Change and Controller Settings

Adjusting air change and initial settings for the RD3 program controller.

Failure Behavior and Safety Devices

Temperature Safety Devices (Class 2 & 3.1)

Details on the function and operation of temperature safety devices.

Maintenance, Cleaning, and Service

Cleaning and Decontamination

Procedures for cleaning and decontaminating the chamber after use.

Troubleshooting

Heating and Controller Faults

Troubleshooting guide for heating system and controller malfunctions.

Binder FP 115 Specifications

General IconGeneral
Temperature Range5 °C above ambient temperature to 300 °C
Temperature Fluctuation±0.1 °C
Permitted load per shelf30 kg
Internal Dimensions (W x H x D)600 x 480 x 400 mm
Power Supply230 V, 50/60 Hz
TypeForced convection oven
Interior Volume115 liters
Number of shelves standard/max2/