Issue 02/2022 Art. No. 7001-0292
Operating Manual
BD (E3.1)
|
Incubators
Avantgarde.Line with natural convection
BF (E3.1)
|
Incubators
Avantgarde.Line with forced convection
ED (E3.1)
|
Drying and heating ovens
Avantgarde.Line with natural convection
FD (E3.1)
|
Drying and heating ovens
Avantgarde.Line with forced convection
FED (E3.1)
|
Drying and heating ovens
Avantgarde.Line with forced convection
and enhanced timer functions
with microprocessor temperature controller
Model Model version Art. No. Model Model version Art. No.
BD 56
BD056-230V 9010/ 9110-0323
ED 260
ED260-230V 9010/ 9110-0339
BD056UL-120V 9010/ 9110-0324 ED260UL-240V 9010/ 9110-0340
BD 115
BD115-230V 9010/ 9110-0325 ED 720 ED720-400V 9010/ 9110-0341
BD115UL-120V 9010/ 9110-0326
FD 56
FD056-230V 9010/ 9110-0303
BD 260
BD260-230V 9010/ 9110-0329 FD056UL-120V 9010/ 9110-0304
BD260UL-120V 9010/ 9110-0330
FD 115
FD115-230V 9010/ 9110-0305
BD 720
BD720-230V 9010/ 9110-0331 FD115UL-120V 9010/ 9110-0306
BD720UL-240V 9010/ 9110-0332
FD 260
FD260-230V 9010/ 9110-0309
BF 56
BF056-230V 9010/ 9110-0313 FD260UL-240V 9010/ 9110-0310
BF056UL-120V 9010/ 9110-0314 FD 720 FD720-400V 9010/ 9110-0311
BF 115
BF115-230V 9010/ 9110-0315
FED 56
FED056-230V 9010/ 9110-0295
BF115UL-120V 9010/ 9110-0316 FED056UL-120V 9010/ 9110-0296
BF 260
BF260-230V 9010/ 9110-0319
FED 115
FED115-230V 9010/ 9110-0293
BF260UL-120V 9010/ 9110-0320 FED115UL-120V 9010/ 9110-0294
BF 720
BF720-230V 9010/ 9110-0321
FED 260
FED260-230V 9010/ 9110-0299
BF720UL-240V 9010/ 9110-0322 FED260UL-240V 9010/ 9110-0300
ED 56
ED056-230V 9010/ 9110-0333
FED 720
FED720-400V 9010/ 9110-0301
ED056UL-120V 9010/ 9110-0334 FED720UL-208V 9010/ 9110-0302
ED 115
ED115-230V 9010/ 9110-0335
ED115UL-120V 9010/ 9110-0336
BINDER GmbH
Address: Post office box 102, 78502 Tuttlingen, Germany Phone: +49 7462 2005 0
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com
E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555
Service Fax: +49 7462 2005 93 555 Service E-Mail: customerservice@binder-world.com
Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3
Service Hotline Asia Pacific: +852 390 705 04 or +852 390 705 03
Service Hotline Russia and CIS: +7 495 988 15 16

Table of Contents

Other manuals for Binder BF 115

Questions and Answers

  • D
    davidcarlsonAug 20, 2025
    What to do if Binder BF 115 Laboratory Equipment chamber has no function?
    • G
      Gregory NunezAug 20, 2025
      First, check the connection to the power supply. If that's not the issue, the overtemperature protective device class 1 might have turned off the chamber. Disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, further assistance is needed.
  • S
    Spencer WilliamsonAug 22, 2025
    Why is the set-point temperature not reached in my Binder Laboratory Equipment after the specified time?
    • P
      Patrick WareAug 22, 2025
      Several factors could be the reason. Make sure the chamber door is completely closed and that the door gasket isn't defective (replace if needed). The controller may also need calibration and adjustment. Finally, verify the power supply for the correct voltage.
  • M
    Michael FitzpatrickAug 25, 2025
    What to do if Binder BF 115 Laboratory Equipment chamber doesn’t heat up but the controller display is working?
    • J
      James FarmerAug 25, 2025
      The timer might have run off. Try re-programming the timer or turning it off.
  • A
    Angel NguyenAug 26, 2025
    Why is my Binder BF 115 chamber heating permanently and not holding the set-point?
    • J
      jnorrisAug 26, 2025
      The controller may not be adjusted, so calibrate and adjust it. If the issue persists, the controller might be defective and further assistance is needed.
  • M
    melaniewalshAug 27, 2025
    What to do if my ED, FD, or FED Binder Laboratory Equipment chamber has no function and shows the alarm message “!TLim”?
    • L
      Lauren JamesAug 28, 2025
      The safety device class 2 has likely turned off the chamber. Allow the chamber to cool down. Check the temperature set-point and the settings of the safety device class 2. If appropriate, select a suitable limit value. If the issue persists, further assistance is needed.
  • B
    baxtereddieAug 29, 2025
    What to do if my Binder BF 115 chamber has no function?
    • J
      jesse98Aug 29, 2025
      First, check the connection to the power supply. If the overtemperature protective device class 1 has turned off the chamber, disconnect the chamber from the power supply for at least 10 seconds and let it cool down. If the device responds again, further assistance is needed.
  • C
    Cody BoydAug 31, 2025
    What to do if my ED, FD, FED Binder BF 115 Laboratory Equipment chamber has no function and shows alarm message “!TLim”?
    • C
      Crystal DoyleAug 31, 2025
      This alarm indicates that safety device class 2 has turned off the chamber. Allow the chamber to cool down. Check the settings of the temperature set-point and the safety device class 2, and select a suitable limit value if needed.
  • T
    Tamara DrakeSep 2, 2025
    What to do if my Binder BF 115 chamber doesn’t heat up and there is no heating icon in the display, but controller display is working?
    • D
      dhernandezSep 2, 2025
      The timer may have run off. Try re-programming the timer or turn it off.
  • R
    roy15Sep 4, 2025
    What to do if my BD, BF Binder Laboratory Equipment displays alarm message “!TProt”?
    • T
      Tyler WhiteSep 4, 2025
      The alarm message “!TProt” indicates that safety device class 3.1 has responded. Check the settings of the temperature set-point and of the safety device class 3.1.
  • R
    Robert AlvaradoSep 6, 2025
    What to do if my Binder BF 115 chamber is heating permanently, and the set-point is not held?
    • C
      Cynthia CortezSep 6, 2025
      The controller may not be adjusted correctly. Calibrate and adjust the controller.

Summary

SAFETY

Personnel Qualification and Manual Usage

Requirements for qualified personnel and proper use of the operating manual.

CHAMBER DESCRIPTION & INSTALLATION

Electrical Connection

Details on connecting the chamber to the power supply.

OPERATION AND SETTINGS

Controller Interface and Settings

Navigating the controller, setting language, date/time, units, set-points, and passwords.

OVERTEMPERATURE PROTECTION AND DATA MANAGEMENT

Overtemperature Protection Features

Understanding and setting overtemperature safety devices and controllers.

OPTIONS AND MAINTENANCE

Troubleshooting and Service

Simple troubleshooting steps and procedure for sending the chamber for repair.

Binder BF 115 Specifications

General IconGeneral
BrandBinder
ModelBF 115
CategoryLaboratory Equipment
LanguageEnglish